Dynamic perfusion CT Optimizing the temporal resolution for the calculation of perfusion CT parameters in stroke patients

被引:46
作者
Kaemena, Andreas
Streitparth, Florian
Grieser, Christian
Lehrnkuhl, Lukas
Jamil, Basil
Wojtal, Katarzyna
Ricke, Jens
Pech, Maciej
机构
[1] Charite Med Univ Berlin, Dept Radiol, D-13353 Berlin, Germany
[2] Charite Med Univ Berlin, Dept Radiotherapy, D-10117 Berlin, Germany
关键词
perfusion CT (PCT); stroke; apoplexy; cerebral infarction; computed tomography (CT);
D O I
10.1016/j.ejrad.2007.02.025
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the influence of different temporal sampling rates on the accuracy of the results from cerebral perfusion CTs in patients with an acute ischemic stroke. Material and methods: Thirty consecutive patients with acute stroke symptoms received a dynamic perfusion CT (LightSpeed 16, GE). Forty millilitres of iomeprol (Imeron 400) were administered at an injection rate of 4 ml/s. After a scan delay of 7 s, two adjacent 10 mm slices at 80 kV and 190 mA were acquired in a cine mode technique with a cine duration of 49 s. Parametric maps for the blood flow (1317), blood volume (BV) and mean transit time (MTT) were calculated for temporal sampling intervals of 0.5, 1, 2, 3 and 4 s using GE's Perfusion 3 software package. In addition to the quantitative ROI data analysis, a visual perfusion map analysis was performed. Results: The perfusion analysis proved to be technically feasible with all patients. The calculated perfusion values revealed significant differences with regard to the BF, BV and MTT, depending on the employed temporal resolution. The perfusion contrast between ischemic lesions and healthy brain tissue decreased continuously at the lower temporal resolutions. The visual analysis revealed that ischemic lesions were best depicted with sampling intervals of 0.5 and 1 s. Conclusion: We recommend a temporal scan resolution of two images per second for the best detection and depiction of ischemic areas. (c) 2007 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:111 / 118
页数:8
相关论文
共 13 条
[1]   Radiation exposure in multi-slice versus single-slice spiral CT: results of a nationwide survey [J].
Brix, G ;
Nagel, HD ;
Stamm, G ;
Veit, R ;
Lechel, U ;
Griebel, J ;
Galanski, M .
EUROPEAN RADIOLOGY, 2003, 13 (08) :1979-1991
[2]   Update on stroke [J].
Chalela, JA ;
Merino, JG ;
Warach, S .
CURRENT OPINION IN NEUROLOGY, 2004, 17 (04) :447-451
[3]   Intra-arterial prourokinase for acute ischemic stroke - The PROACT II study: A randomized controlled trial [J].
Furlan, A ;
Higashida, R ;
Wechsler, L ;
Gent, M ;
Rowley, H ;
Kase, C ;
Pessin, M ;
Ahuja, A ;
Callahan, F ;
Clark, WM ;
Silver, F ;
Rivera, F .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1999, 282 (21) :2003-2011
[4]  
Hirata Masaaki, 2005, Radiat Med, V23, P97
[5]   Cerebral perfusion CT: Technique and clinical applications [J].
Hoeffner, EG ;
Case, I ;
Jain, R ;
Gujar, SK ;
Shah, GV ;
Deveikis, JP ;
Carlos, RC ;
Thompson, BG ;
Harrigan, MR ;
Mukherji, SK .
RADIOLOGY, 2004, 231 (03) :632-644
[6]   Perfusion CT of the brain: Diagnostic approach for early detection of ischemic stroke [J].
Koenig, M ;
Klotz, E ;
Luka, B ;
Venderink, DJ ;
Spittler, JF ;
Heuser, L .
RADIOLOGY, 1998, 209 (01) :85-93
[7]   TISSUE-PLASMINOGEN ACTIVATOR FOR ACUTE ISCHEMIC STROKE [J].
MARLER, JR ;
BROTT, T ;
BRODERICK, J ;
KOTHARI, R ;
ODONOGHUE, M ;
BARSAN, W ;
TOMSICK, T ;
SPILKER, J ;
MILLER, R ;
SAUERBECK, L ;
JARRELL, J ;
KELLY, J ;
PERKINS, T ;
MCDONALD, T ;
RORICK, M ;
HICKEY, C ;
ARMITAGE, J ;
PERRY, C ;
THALINGER, K ;
RHUDE, R ;
SCHILL, J ;
BECKER, PS ;
HEATH, RS ;
ADAMS, D ;
REED, R ;
KLEI, M ;
HUGHES, S ;
ANTHONY, J ;
BAUDENDISTEL, D ;
ZADICOFF, C ;
RYMER, M ;
BETTINGER, I ;
LAUBINGER, P ;
SCHMERLER, M ;
MEIROSE, G ;
LYDEN, P ;
RAPP, K ;
BABCOCK, T ;
DAUM, P ;
PERSONA, D ;
BRODY, M ;
JACKSON, C ;
LEWIS, S ;
LISS, J ;
MAHDAVI, Z ;
ROTHROCK, J ;
TOM, T ;
ZWEIFLER, R ;
DUNFORD, J ;
ZIVIN, J .
NEW ENGLAND JOURNAL OF MEDICINE, 1995, 333 (24) :1581-1587
[8]   Perfusion CT: a worthwhile enhancement? [J].
Miles, KA ;
Griffiths, MR .
BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (904) :220-231
[9]  
Roberts HC, 2001, AM J NEURORADIOL, V22, P1077
[10]  
Wintermark M, 2000, AM J NEURORADIOL, V21, P1881