Use of pseudo-first order kinetics in flow injection for determination of trace inorganic iodine

被引:20
作者
Choengchan, N
Lukkanakul, K
Ratanawimarnwong, N
Waiyawat, W
Wilairata, P
Nacapricha, D
机构
[1] Mahidol Univ, Fac Sci, Dept Chem, Bangkok 10400, Thailand
[2] Mahidol Univ, Inst Sci & Technol Res & Dev, Salaya 73170, Nakorn Pathom, Thailand
关键词
iodate; iodide; kinetic determination; drinking water; flow injection;
D O I
10.1016/j.aca.2003.08.062
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work utilizes pseudo-first order kinetics for the determination of iodine at the part per billion (ppb) levels. Flow injection (FI) system was developed and optimised for conditions in which the indicator reaction between Ce(IV) and As(III) gives first order kinetics. This oxidation-reduction reaction is catalysed by iodide ion and has been used for determination of trace levels of iodide for many years. Under pseudo-first order condition, analysis can be made in two ways, either a full kinetic or a fixed-time method. System development is presented as well as the characteristic evaluation of the kinetics. The system, with a simple and interferent-free in-line reduction step, has allowed successful quantitation of iodate, which has never been reported for this kinetic method. The system was used to determine total inorganic iodine in bottled drinking water that had been supplemented with iodate ion to the level of 100 mug Il(-1). (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:115 / 122
页数:8
相关论文
共 17 条
[1]   RAPID METHOD FOR THE DIRECT DETERMINATION OF INORGANIC IODINE IN PLASMA USING ION-EXCHANGE CHROMATOGRAPHY AND THE SANDELL AND KOLTHOFF REACTION [J].
AUMONT, G ;
TRESSOL, JC .
ANALYST, 1987, 112 (06) :875-878
[2]  
Chantore W, 1999, LAB ROBOTICS AUTOMAT, V11, P37, DOI 10.1002/(SICI)1098-2728(1999)11:1<37::AID-LRA5>3.0.CO
[3]  
2-5
[4]   2 SIMPLE METHODS FOR MEASURING IODINE IN URINE [J].
DUNN, JT ;
CRUTCHFIELD, HE ;
GUTEKUNST, R ;
DUNN, AD .
THYROID, 1993, 3 (02) :119-123
[5]  
Erustes JA, 2001, J AOAC INT, V84, P337
[6]   DETERMINATION OF SUBMICROGRAM AMOUNTS OF IODIDE BY ITS CATALYTIC EFFECT ON SUBSTITUTION REACTION OF MERCURY(II)-PAR COMPLEX WITH CYDTA [J].
FUNAHASH.S ;
TABATA, M ;
TANAKA, M .
ANALYTICA CHIMICA ACTA, 1971, 57 (02) :311-&
[7]   STOPPED-FLOW DETERMINATION OF IODIDE IN PHARMACEUTICAL AND FOOD SAMPLES [J].
GUTIERREZ, MC ;
GOMEZHENS, A ;
PEREZBENDITO, D .
ANALYST, 1989, 114 (01) :89-92
[8]  
MICHAEL GH, 1968, MODERN NUTR HLTH DIS, P372
[9]  
MILLER JC, 2000, STAT CHEMOMETRICS AN, P48
[10]   Continuous and stopped flow injection for catalytic determination of total iodine in urine [J].
Nacapricha, D ;
Muangkaew, S ;
Ratanawimarnwong, N ;
Shiowatana, J ;
Grudpan, K .
ANALYST, 2001, 126 (01) :121-126