Collapse of a stent-graft following treatment of a traumatic thoracic aortic rupture

被引:96
作者
Idu, MM
Reekers, JA
Balm, R
Ponsen, KJ
de Mol, BAJM
Legemate, DA
机构
[1] Acad Med Ctr, Dept Vasc Surg, NL-1100 DD Amsterdam, Netherlands
[2] Acad Med Ctr, Dept Radiol, NL-1100 DD Amsterdam, Netherlands
[3] Acad Med Ctr, Trauma Unit, NL-1100 DD Amsterdam, Netherlands
[4] Acad Med Ctr, Dept Cardiothorac Surg, NL-1100 DD Amsterdam, Netherlands
关键词
thoracic aorta; trauma; aortic rupture; endovascular repair; stent-graft; complication; stent-graft collapse;
D O I
10.1583/04-1515R.1
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To report a collapsed stent-graft used to treat a traumatic aortic rupture. Case Report: A Gore TAG stent-graftwas placed in a 20-year-old man with multiple injuries. Postimplantation computed tomographic angiography (CTA) demonstrated no contrast extravasation and total exclusion of the traumatic rupture. Routine CTA 3 months after implantation revealed a collapsed stent-graft located in the outer curve of the distal aortic arch. A Talent stent-graft was placed successfully within the collapsed prosthesis. Postimplantation CTA demonstrated no contrast extravasation and good apposition of the endograft to the aortic wall. At 6 months, the repair remains secure; there is no sign of graft collapse or endoleak. Conclusions: Collapse of stent-grafts can occur after treatment for traumatic aortic ruptures; endovascular methods can be used to restore a satisfactory luminal contour.
引用
收藏
页码:503 / 507
页数:5
相关论文
共 9 条
  • [1] Surgical versus endovascular treatment of traumatic thoracic aortic rupture
    Amabile, P
    Collart, F
    Gariboldi, V
    Rollet, G
    Bartoli, JA
    Piquet, P
    [J]. JOURNAL OF VASCULAR SURGERY, 2004, 40 (05) : 873 - 879
  • [2] BALM R, 2004, VASCULAR ENDOVASCULA, P226
  • [3] Angiographic description of blunt traumatic injuries to the thoracic aorta with specific relevance to endograft repair
    Borsa, JJ
    Hoffer, EK
    Karmy-Jones, R
    Fontaine, AB
    Bloch, RD
    Yoon, JK
    So, CR
    Meissner, MH
    Demirer, S
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2002, 9 : 84 - 91
  • [4] Endovascular stent grafts for acute blunt aortic injury
    Dunham, MB
    Zygun, D
    Petrasek, P
    Kortbeek, JB
    Karmy-Jones, R
    Moore, RD
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2004, 56 (06): : 1173 - 1178
  • [5] Görich J, 2002, J ENDOVASC THER, V9, P39
  • [6] Disappointing results with a new commercially available thoracic endograft
    Melissano, G
    Tshomba, Y
    Civilini, E
    Chiesa, R
    [J]. JOURNAL OF VASCULAR SURGERY, 2004, 39 (01) : 124 - 130
  • [7] Emergency stent-graft placement for hemorrhage control in acute thoracic aortic rupture
    Melnitchouk, S
    Pfammatter, T
    Kadner, A
    Dave, H
    Witzke, H
    Trentz, O
    Turina, M
    Lachat, M
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2004, 25 (06) : 1032 - 1038
  • [8] Emergency endovascular treatment of thoracic aortic rupture in three accident victims with multiple injuries
    Nio, D
    Vos, PM
    de Mol, BAJM
    Ponsen, IJ
    Reekers, JA
    Balm, R
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2002, 9 : 60 - 66
  • [9] Endoluminal stent-graft placement for acute rupture of the descending thoracic aorta
    Scheinert, D
    Krankenberg, H
    Schmidt, A
    Gummert, JF
    Nitzsche, S
    Scheinert, S
    Bräunlich, S
    Sorge, I
    Krakor, R
    Biamino, G
    Schuler, G
    Mohr, FW
    [J]. EUROPEAN HEART JOURNAL, 2004, 25 (08) : 694 - 700