Attenuation effects of biliary endoprostheses on therapeutic radiation

被引:4
作者
MayoSmith, WW
Dawson, SL
Mauceri, T
Mueller, PR
机构
[1] HARVARD UNIV, MASSACHUSETTS GEN HOSP, SCH MED, DEPT ABDOMINAL & INTERVENT RADIOL, BOSTON, MA USA
[2] HARVARD UNIV, MASSACHUSETTS GEN HOSP, SCH MED, DEPT RADIAT ONCOL, BOSTON, MA USA
关键词
bile ducts; stents and prostheses; stenosis or obstruction; radiations; measurement; therapeutic radiology;
D O I
10.1148/radiology.199.2.8668815
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Seven commercially available biliary stents were tested to determine if they caused any clinically important attenuation of therapeutic radiation. Attenuation effects were evaluated in models of brachytherapy (iridium-192 source) and external beam radiation therapy (cobalt-60 source), with use of parameters that mimicked those used in patient treatment for biliary neoplasms. No stent demonstrated clinically important attenuation or scatter of therapeutic radiation, so local dosimetry was not affected.
引用
收藏
页码:571 / 572
页数:2