Structural Integrity of Balloon-Expandable Stents After Transcatheter Aortic Valve Replacement Assessment by Multidetector Computed Tomography

被引:54
作者
Willson, Alexander B. [1 ]
Webb, John G. [1 ]
Gurvitch, Ronen [1 ]
Wood, David A. [1 ]
Toggweiler, Stefan [1 ]
Binder, Ronald [1 ]
Freeman, Melanie [1 ]
Madden, Mark [1 ]
Hague, Cameron [1 ]
Leipsic, Jonathon [1 ]
机构
[1] Univ British Columbia, St Pauls Hosp, Vancouver, BC V6Z 1Y6, Canada
关键词
aortic stenosis; multidetector computed tomography; transcatheter aortic valve implantation; transcatheter aortic valve replacement; IMPLANTATION; REOPERATION; DEPLOYMENT; OUTCOMES;
D O I
10.1016/j.jcin.2012.03.007
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
Objectives This study sought to evaluate the structural integrity of balloon-expandable stents used in transcatheter aortic valve replacement. Background Underexpansion, deformation, or fracture of stent frames may affect transcatheter heart valve (THV) function and durability. Methods Patients >1 year after transcatheter aortic valve replacement underwent multidetector computed tomography. Geometry of the stent frame was assessed for circularity; eccentricity; minimum and maximum external diameter; and expansion at the inflow, mid-stent, and outflow levels, as well as for stent fracture. THV noncircularity was defined as stent eccentricity >10% (1 - minimum diameter/maximum diameter) and THV underexpansion when expansion <90% (multidetector computed tomography derived external valve area/nominal external valve area). Echocardiography was performed after implantation and annually. Results Fifty patients underwent multidetector computed tomography at an average of 2.5 +/- 0.9 years after transcatheter aortic valve replacement (35 Sapien, 8 Sapien XT, and 7 Cribier-Edwards valves [all Edwards Lifesciences, Irvine, California). The mean external diameter for the 23- and 26-mm valves was 23.3 +/- 0.9 mm and 25.9 +/- 0.9 mm, respectively. Circularity was present in 96% (48 of 50) and median eccentricity was 2.0% (interquartile range: 1.2% to 3.0%). Mean THV expansion was 104.1 +/- 7.4%, which increased from stent inflow to outflow (100.8 +/- 7.6% vs. 108.1 +/- 6.9%, p < 0.001). Stent fracture was not observed. Underexpanded valves (8% [4 of 50]) and noncircular valves (4% [2 of 50]) demonstrated stable hemodynamic function on annual echocardiography. Conclusions Balloon-expandable aortic valves have excellent rates of circularity with low eccentricity and maintain full expansion without stent fracture at an average 2.5 years after implantation. (J Am Coll Cardiol Intv 2012;5:525-32) (C) 2012 by the American College of Cardiology Foundation
引用
收藏
页码:525 / 532
页数:8
相关论文
共 17 条
[1]
Bongartz G, 2004, APPENDIX C EUROPEAN
[2]
2-Year Follow-Up of Patients Undergoing Transcatheter Aortic Valve Implantation Using a Self-Expanding Valve Prosthesis [J].
Buellesfeld, Lutz ;
Gerckens, Ulrich ;
Schuler, Gerhard ;
Bonan, Raoul ;
Kovac, Jan ;
Serruys, Patrick W. ;
Labinaz, Marino ;
den Heijer, Peter ;
Mullen, Michael ;
Tymchak, Wayne ;
Windecker, Stephan ;
Mueller, Ralf ;
Grube, Eberhard .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (16) :1650-1657
[3]
Transcatheter aortic valve implantation: role of multi-detector row computed tomography to evaluate prosthesis positioning and deployment in relation to valve function [J].
Delgado, Victoria ;
Ng, Arnold C. T. ;
van de Veire, Nico R. ;
van der Kley, Frank ;
Schuijf, Joanne D. ;
Tops, Laurens F. ;
de Weger, Arend ;
Tavilla, Giuseppe ;
de Roos, Albert ;
Kroft, Lucia J. ;
Schalij, Martin J. ;
Bax, Jeroen J. .
EUROPEAN HEART JOURNAL, 2010, 31 (09) :1114-1123
[4]
Reoperation for prosthetic aortic valve obstruction in the era of echocardiography: Trends in diagnostic testing and comparison with surgical findings [J].
Girard, SE ;
Miller, FA ;
Orszulak, TA ;
Mullany, CJ ;
Montgomery, S ;
Edwards, WD ;
Tazelaar, HD ;
Malouf, JF ;
Tajik, AJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (02) :579-584
[5]
Transcatheter Aortic Valve Implantation Durability of Clinical and Hemodynamic Outcomes Beyond 3 Years in a Large Patient Cohort [J].
Gurvitch, R. ;
Wood, D. A. ;
Tay, E. L. ;
Leipsic, J. ;
Ye, J. ;
Lichtenstein, S. V. ;
Thompson, C. R. ;
Carere, R. G. ;
Wijesinghe, N. ;
Nietlispach, F. ;
Boone, R. H. ;
Lauck, S. ;
Cheung, A. ;
Webb, J. G. .
CIRCULATION, 2010, 122 (13) :1319-1327
[6]
Multidetector Computed Tomography in Transcatheter Aortic Valve Implantation [J].
Leipsic, Jonathon ;
Gurvitch, Ronen ;
LaBounty, Troy M. ;
Min, James K. ;
Wood, David ;
Johnson, Mark ;
Ajlan, Amr M. ;
Wijesinghe, Namal ;
Webb, John G. .
JACC-CARDIOVASCULAR IMAGING, 2011, 4 (04) :416-429
[7]
Standardized Endpoint Definitions for Transcatheter Aortic Valve Implantation Clinical Trials A Consensus Report From the Valve Academic Research Consortium [J].
Leon, Martin B. ;
Piazza, Nicolo ;
Nikolsky, Eugenia ;
Blackstone, Eugene H. ;
Cutlip, Donald E. ;
Kappetein, Arie Pieter ;
Krucoff, Mitchell W. ;
Mack, Michael ;
Mehran, Roxana ;
Miller, Craig ;
Morel, Marie-angele ;
Petersen, John ;
Popma, Jeffrey J. ;
Takkenberg, Johanna J. M. ;
Vahanian, Alec ;
van Es, Gerrit-Anne ;
Vranckx, Pascal ;
Webb, John G. ;
Windecker, Stephan ;
Serruys, Patrick W. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (03) :253-269
[8]
Structural deterioration of the Freestyle aortic valve: Mode of presentation and mechanisms [J].
Mohammadi, Siamak ;
Baillot, Richard ;
Voisine, Pierre ;
Mathieu, Patrick ;
Dagenais, Francois .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2006, 132 (02) :401-406
[9]
Current Balloon-Expandable Transcatheter Heart Valve and Delivery Systems [J].
Nietlispach, Fabian ;
Wijesinghe, Namal ;
Wood, David ;
Carere, Ronald G. ;
Webb, John G. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 75 (02) :295-300
[10]
A clinical protocol for analysis of the structural integrity of the Medtronic CoreValve System® frame and its application in patients with 1-year minimum follow-up [J].
Piazza, Nicolo ;
Grube, Eberhard ;
Gerckens, Ulrich ;
Schuler, Gerhard ;
Linke, Axel ;
den Heijer, Peter ;
Kovacs, Jan ;
Spyt, Tom ;
Laborde, Jean-Claude ;
Morel, Marie-Angele ;
Nuis, Rutger-Jan ;
Garcia-Garcia, Hector M. ;
de Jaegere, Peter ;
Serruys, Patrick W. .
EUROINTERVENTION, 2010, 5 (06) :680-686