Biocompatibility of the Ex-PRESS miniature glaucoma drainage implant

被引:111
作者
Nyska, A
Glovinsky, Y
Belkin, M
Epstein, Y
机构
[1] NIEHS, Lab Expt Pathol, Res Triangle Pk, NC 27709 USA
[2] Tel Aviv Univ, Goldshleger Eye Res Inst, Tel Hashomer, Israel
[3] Tel Aviv Univ, Sackler Fac Med, IL-69978 Tel Aviv, Israel
关键词
aqueous shunts; biocompatibility; drainage device; glaucoma;
D O I
10.1097/00061198-200306000-00017
中图分类号
R77 [眼科学];
学科分类号
100212 [眼科学];
摘要
Purpose: Based on lessons learned from earlier attempts, a novel miniature glaucoma implant, Ex-PRESS, was developed in 1998. The current study summarizes the histopathologic evaluation of this device implanted in the eyes of rabbits. Methods: The device was implanted into the anterior chamber at the corneoscleral junction in 1 eye each of 8 white New Zealand rabbits, while the contralateral eye served as control. Three and 6 months after implantation, the rabbits were killed and their eyes were enucleated and processed histologically, leaving the device in situ when sectioning. Results: Three and 6 months postoperatively, the local tissue reaction typically consisted of an enveloping, thin, mature, fibrotic capsule (thickness <0.04 mm), devoid of inflammatory cells. This capsule surrounded approximately 25% of the implant surface area present in the sections. The lumina of the devices were devoid of inflammatory exudates or other obstructions in all specimens examined, suggesting free flow of fluid. Conclusions: The implantation of the Ex-PRESS miniature glaucoma shunt resulted in minimal capsular reaction. Considering the high reactivity of the rabbit eye, it is possible that this implant will induce a smaller cellular inflammatory reaction in the human eye.
引用
收藏
页码:275 / 280
页数:6
相关论文
共 24 条
[1]
[Anonymous], 1986, HDB BIOMATERIAL EVAL
[2]
Comparison of different biomaterials for glaucoma drainage devices - Part 2 [J].
Ayyala, RS ;
Michelini-Norris, B ;
Flores, A ;
Haller, E ;
Margo, CE .
ARCHIVES OF OPHTHALMOLOGY, 2000, 118 (08) :1081-1084
[3]
Ayyala RS, 1999, ARCH OPHTHALMOL-CHIC, V117, P233
[4]
TRABECULECTOMY WITH SIMULTANEOUS TOPICAL APPLICATION OF MITOMYCIN-C IN REFRACTORY GLAUCOMA [J].
CHEN, CW ;
HUANG, HT ;
BAIR, JS ;
LEE, CC .
JOURNAL OF OCULAR PHARMACOLOGY, 1990, 6 (03) :175-182
[5]
GROSSBLATT N, 1996, GUIDE CARE USE LAB A
[6]
THE MECHANISMS OF PASSIVE DISSOLUTION OF TITANIUM IN A MODEL PHYSIOLOGICAL ENVIRONMENT [J].
HEALY, KE ;
DUCHEYNE, P .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (03) :319-338
[7]
HEUER DK, 1984, OPHTHALMOLOGY, V91, P384
[8]
Jallet V, 1999, J BIOMED MATER RES, V47, P260, DOI 10.1002/(SICI)1097-4636(199911)47:2<260::AID-JBM17>3.0.CO
[9]
2-J
[10]
Kaplan-Messas A, 2001, INVEST OPHTH VIS SCI, V42, pS552