A diode-laser sensor has been developed for sensitive measurements of CO2, H2O, N2O, and NH3 concentrations in various flowfields using absorption spectroscopy and extractive sampling techniques. An external-cavity diode laser with a tuning range of 1.953-2.057 mu m was used to record absorption lineshapes from measured transitions in the CO2 4 nu(2)(0) + nu(3), nu(1 +) 2 nu(2)(0) + nu(3), and 2 nu(1) + nu(3) bands, H2O + nu(3) and nu(1) + nu(2) bands, N2O 2 nu(1) + 4(2)(0), nu(2)(1) + 2 nu(3), 3 nu(1) + 2 nu(2)(0), and 4 nu(1) bands, and NH3 nu(1) + nu(4) and nu(1) + nu(4) bands. Measured CO2, H2O, and N2O survey spectra were compared to calculations to verify the HITRAN96 database and used to determine optimum transitions for species detection. Individual lineshape measurements were used to determine fundamental spectroscopic parameters including the line strength, line-center frequency, and self-broadening coefficient of the probed transition. The results represent the first measurements of CO2, H2O, N2O, and NH3 absorption near 2.0 mu m using room-temperature near-IR diode lasers.
引用
收藏
页码:283 / 288
页数:6
相关论文
共 14 条
  • [1] VIBRATION-ROTATION BANDS OF N-14(2) O-16 - 1.2 MICRON-3.3 MICRON REGION
    AMIOT, C
    GUELACHVILI, G
    [J]. JOURNAL OF MOLECULAR SPECTROSCOPY, 1974, 51 (03) : 475 - 491
  • [2] MULTIPLEXED DIODE-LASER SENSOR SYSTEM FOR SIMULTANEOUS H2O, O-2, AND TEMPERATURE-MEASUREMENTS
    BAER, DS
    HANSON, RK
    NEWFIELD, ME
    GOPAUL, NKJM
    [J]. OPTICS LETTERS, 1994, 19 (22) : 1900 - 1902
  • [3] Diode laser absorption measurements of CH3Cl and CH4 near 1.65 mu m
    Chou, SI
    Baer, DS
    Hanson, RK
    [J]. APPLIED OPTICS, 1997, 36 (15): : 3288 - 3293
  • [4] CALCULATED TABULATIONS OF H2O LINE BROADENING BY H2O,N2,O2, AND CO2 AT HIGH-TEMPERATURE
    DELAYE, C
    HARTMANN, JM
    TAINE, J
    [J]. APPLIED OPTICS, 1989, 28 (23): : 5080 - 5087
  • [5] PARAMETRIC DECONVOLUTION METHOD - APPLICATION TO 2 BANDS OF N2O IN 1.9-MU-M REGION
    LACOME, N
    LEVY, A
    [J]. JOURNAL OF MOLECULAR SPECTROSCOPY, 1978, 71 (1-3) : 175 - 192
  • [6] NOVEL GEOMETRY FOR SINGLE-MODE SCANNING OF TUNABLE LASERS
    LIU, K
    LITTMAN, MG
    [J]. OPTICS LETTERS, 1981, 6 (03) : 117 - 118
  • [7] SYNCHRONOUS CAVITY MODE AND FEEDBACK WAVELENGTH SCANNING IN DYE-LASER OSCILLATORS WITH GRATINGS
    MCNICHOLL, P
    METCALF, HJ
    [J]. APPLIED OPTICS, 1985, 24 (17): : 2757 - 2761
  • [8] A diode-laser absorption sensor system for combustion emission measurements
    Mihalcea, RM
    Baer, DS
    Hanson, RK
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 1998, 9 (03) : 327 - 338
  • [9] Diode laser sensor for measurements of CO, CO2, and CH4 in combustion flows
    Mihalcea, RM
    Baer, DS
    Hanson, RK
    [J]. APPLIED OPTICS, 1997, 36 (33): : 8745 - 8752
  • [10] THE HITRAN MOLECULAR DATABASE - EDITIONS OF 1991 AND 1992
    ROTHMAN, LS
    GAMACHE, RR
    TIPPING, RH
    RINSLAND, CP
    SMITH, MAH
    BENNER, DC
    DEVI, VM
    FLAUD, JM
    CAMYPEYRET, C
    PERRIN, A
    GOLDMAN, A
    MASSIE, ST
    BROWN, LR
    TOTH, RA
    [J]. JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 1992, 48 (5-6) : 469 - 507