High-performance liquid chromatographic separation of microcystins and nodularin, cyanobacterial peptide toxins, on C18 and amide C16 sorbents

被引:56
作者
Spoof, L [1 ]
Karlsson, K [1 ]
Meriluoto, J [1 ]
机构
[1] Abo Akad Univ, Dept Biochem & Pharm, Turku 20521, Finland
基金
芬兰科学院;
关键词
stationary phases; LC; mobile phase composition; microcystins; toxins; nodularin;
D O I
10.1016/S0021-9673(00)01099-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Four C-18 columns and a novel amide C-16 column were assessed in the HPLC separation of eight microcystins and nodularin-R. Gradient mobile phases of acetonitrile combined with trifluoroacetic acid, formic acid or ammonium acetate were compared. Special attention was paid to the resolution of four possible coeluting microcystin pairs. Generally speaking, the acidic mobile phases were superior to the ammonium acetate-based mobile phase in terms of resolution and selectivity. The amide C-16 column had the best overall performance and unique selectivity properties. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:225 / 236
页数:12
相关论文
共 23 条
[1]   Laboratory-scale purification of microcystins using flash chromatography and reversed-phase high-performance liquid chromatography [J].
Edwards, C ;
Lawton, LA ;
Coyle, SM ;
Ross, P .
JOURNAL OF CHROMATOGRAPHY A, 1996, 734 (01) :163-173
[2]   LIQUID-CHROMATOGRAPHIC DETERMINATION OF THE CYANOGINOSINS, TOXINS PRODUCED BY THE CYANOBACTERIUM MICROCYSTIS-AERUGINOSA [J].
GATHERCOLE, PS ;
THIEL, PG .
JOURNAL OF CHROMATOGRAPHY, 1987, 408 :435-440
[3]   PREDICTION OF PEPTIDE RETENTION TIMES IN REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY .1. DETERMINATION OF RETENTION COEFFICIENTS OF AMINO-ACID-RESIDUES OF MODEL SYNTHETIC PEPTIDES [J].
GUO, DC ;
MANT, CT ;
TANEJA, AK ;
PARKER, JMR ;
HODGES, RS .
JOURNAL OF CHROMATOGRAPHY, 1986, 359 :499-517
[4]   PREDICTION OF PEPTIDE RETENTION TIMES IN REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY .2. CORRELATION OF OBSERVED AND PREDICTED PEPTIDE RETENTION TIMES AND FACTORS INFLUENCING THE RETENTION TIMES OF PEPTIDES [J].
GUO, DC ;
MANT, CT ;
TANEJA, AK ;
HODGES, RS .
JOURNAL OF CHROMATOGRAPHY, 1986, 359 :519-532
[5]   ISOLATION AND CHARACTERIZATION OF THE MINOR COMPONENTS ASSOCIATED WITH MICROCYSTINS LR AND RR IN THE CYANOBACTERIUM (BLUE-GREEN-ALGAE) [J].
HARADA, K ;
MATSUURA, K ;
SUZUKI, M ;
WATANABE, MF ;
OISHI, S ;
DAHLEM, AM ;
BEASLEY, VR ;
CARMICHAEL, WW .
TOXICON, 1990, 28 (01) :55-64
[6]   STRUCTURAL DETERMINATION OF GEOMETRICAL-ISOMERS OF MICROCYSTINS LR AND RR FROM CYANOBACTERIA BY 2-DIMENSIONAL NMR SPECTROSCOPIC TECHNIQUES [J].
HARADA, K ;
OGAWA, K ;
MATSUURA, K ;
MURATA, H ;
SUZUKI, M ;
WATANABE, MF ;
ITEZONO, Y ;
NAKAYAMA, N .
CHEMICAL RESEARCH IN TOXICOLOGY, 1990, 3 (05) :473-481
[7]  
Harada K., 1996, TOXIC MICROCYSTIS, P103
[8]  
Harada Ken-Ichi, 1997, Natural Toxins, V5, P201, DOI 10.1002/nt.5
[9]   PERSISTENCE OF CYCLIC PEPTIDE TOXINS IN DRIED MICROCYSTIS-AERUGINOSA CRUSTS FROM LAKE MOKOAN, AUSTRALIA [J].
JONES, GJ ;
FALCONER, IR ;
WILKINS, RM .
ENVIRONMENTAL TOXICOLOGY AND WATER QUALITY, 1995, 10 (01) :19-24
[10]   EXTRACTION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF MICROCYSTINS IN RAW AND TREATED WATERS [J].
LAWTON, LA ;
EDWARDS, C ;
CODD, GA .
ANALYST, 1994, 119 (07) :1525-1530