Differences of bone bonding ability and degradation behaviour in vivo between amorphous calcium phosphate and highly crystalline hydroxyapatite coating

被引:197
作者
Nagano, M
Nakamura, T
Kokubo, T
Tanahashi, M
Ogawa, M
机构
[1] KYOTO UNIV, FAC MED, DEPT ORTHOPED SURG, SAKYO KU, KYOTO 60601, JAPAN
[2] KYOTO UNIV, FAC ENGN, DIV MAT CHEM, SAKYO KU, KYOTO 60601, JAPAN
[3] JAPAN SHERWOOD CO LTD, FUKUROI, SHIZUOKA 437, JAPAN
关键词
amorphous calcium phosphate; highly crystalline hydroxyapatite; coating degradation; biomimetic; in vivo; osteoconductivity;
D O I
10.1016/0142-9612(95)00357-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Three types of calcium phosphate coating were formed on polyethersulphone (PES) rectangular plates using a biomimetic method: a 20 mu m thick amorphous calcium phosphate (ACP 20) coating, a 50 mu m thick amorphous calcium phosphate (ACP 50) coating, and a 50 mu m thick highly crystalline hydroxyapatite (hHA 50) coating. Uncoated PES plates were used as a control group. These materials were implanted in the tibiae of rabbits and subcutaneously in rats, and the samples were harvested 8 and 16 weeks thereafter, and were examined histologically. The tensile failure loads at the bone-implant interfaces were determined using the detaching test. Each ACP coating was more degradable than the hHA 50 coating. However, newly formed bone came into direct contact with underlying materials as the coating degraded. No coating degraded in subcutaneous tissue. Soft tissue intervening was seen in uncoated samples. Failure load of ACP 20-, ACP 50- and hHA 50-coated samples were all relatively higher than that of the uncoated samples at each period. Significant increase of failure load was seen in hHA 50-coated samples by 16 weeks, however, no increase was seen in either the uncoated or ACP-coated samples. If coating longevity is desired, then the hHA coating is preferable. However, if only the osteoconducive property of calcium phosphate coating is desired for initial fixation of porous materials, the ACP coating may be advantageous. (C) 1996 Elsevier Science Limited.
引用
收藏
页码:1771 / 1777
页数:7
相关论文
共 28 条
[1]   MECHANISMS AND STRUCTURE OF THE BOND BETWEEN BONE AND HYDROXYAPATITE CERAMICS [J].
BAGAMBISA, FB ;
JOOS, U ;
SCHILLI, W .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (08) :1047-1055
[2]  
BOONE PS, 1989, J BIOMED MATER RES-A, V23, P183
[3]  
BURRI C, 1989, Aktuelle Traumatologie, V19, P297
[4]   INFLUENCE OF SURFACE CHARACTERISTICS ON BONE INTEGRATION OF TITANIUM IMPLANTS - A HISTOMORPHOMETRIC STUDY IN MINIATURE PIGS [J].
BUSER, D ;
SCHENK, RK ;
STEINEMANN, S ;
FIORELLINI, JP ;
FOX, CH ;
STICH, H .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (07) :889-902
[5]   HYDROXYLAPATITE COATING OF POROUS IMPLANTS IMPROVES BONE INGROWTH AND INTERFACE ATTACHMENT STRENGTH [J].
COOK, SD ;
THOMAS, KA ;
DALTON, JE ;
VOLKMAN, TK ;
WHITECLOUD, TS ;
KAY, JF .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (08) :989-1001
[6]  
DEBRUIJN JD, 1992, BIOCERAMICS, V5, P291
[7]   PLASMA SPRAYED COATINGS OF HYDROXYLAPATITE [J].
DEGROOT, K ;
GEESINK, R ;
KLEIN, CPAT ;
SEREKIAN, P .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1987, 21 (12) :1375-1381
[8]   A HISTOLOGICAL AND HISTOMORPHOMETRICAL INVESTIGATION OF FLUORAPATITE, MAGNESIUMWHITLOCKITE, AND HYDROXYLAPATITE PLASMA-SPRAYED COATINGS IN GOATS [J].
DHERT, WJA ;
KLEIN, CPAT ;
JANSEN, JA ;
VANDERVELDE, EA ;
VRIESDE, RC ;
ROZING, PM ;
DEGROOT, K .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (01) :127-138
[9]  
DUCHEYNE P, 1992, CLIN ORTHOP RELAT R, P102
[10]   BONDING OF BONE TO APATITE-COATED IMPLANTS [J].
GEESINK, RGT ;
DEGROOT, K ;
KLEIN, CPAT .