Neurothrombectomy Devices for the Treatment of Acute Ischemic Stroke: State of the Evidence

被引:59
作者
Baker, William L. [1 ]
Colby, Jennifer A. [1 ]
Tongbram, Vanita [1 ]
Talati, Ripple [1 ]
Silverman, Isaac E. [1 ]
White, C. Michael [1 ]
Kluger, Jeffrey [1 ]
Coleman, Craig I. [1 ]
机构
[1] Univ Connecticut, Hartford Hosp, Evidence Based Practice Ctr, Hartford, CT 06102 USA
基金
美国医疗保健研究与质量局;
关键词
MERCI CLOT RETRIEVER; MECHANICAL THROMBECTOMY; INTERVENTIONAL MANAGEMENT; PENUMBRA SYSTEM; POOLED ANALYSIS; AB DEVICE; RECANALIZATION; REVASCULARIZATION; OCCLUSION; THROMBOLYSIS;
D O I
10.7326/0003-4819-154-4-201102150-00306
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Acute ischemic strokes are associated with poor outcomes and high health care burden. Evidence exists evaluating the use of neurothrombectomy devices in patients receiving currently recommended treatments that may have limited efficacy. Purpose: To describe the state of the evidence supporting use of neurothrombectomy devices in the treatment of acute ischemic stroke. Data Sources: MEDLINE, SCOPUS, the Cochrane Central Register of Controlled Trials, the Cochrane Database of Systematic Reviews, and Web of Science were searched, without language restrictions, from their inception through May 2010. The MEDLINE and Cochrane Central Register of Controlled Trials searches were updated through November 2010. Study Selection: Two independent investigators screened citations for human studies of any design or case series or case reports of patients with an acute ischemic stroke that evaluated a neurothrombectomy device and reported at least 1 clinical effectiveness outcome or harm. Data Extraction: Using standardized protocols, 2 independent investigators extracted information about study characteristics and outcomes, and a third reviewer resolved disagreement. Data Synthesis: 87 articles met eligibility criteria, including 18 prospective single-group studies, 7 noncomparative retrospective studies, and 62 case series or case reports. Two U. S. Food and Drug Administration (FDA)-cleared devices, the MERCI Retriever (Concentric Medical, Mountain View, California) (40%) and the Penumbra System (Penumbra, Alameda, California) (9%), represented a large portion of the available data. All prospective and retrospective studies provided data on successful recanalization with widely varying rates (43% to 78% with the MERCI Retriever and 83% to 100% with the Penumbra System). Rates of harms, including symptomatic (16 studies; 0% to 10% with the MERCI Retriever and 0% to 11% with the Penumbra System) or asymptomatic (13 studies; 28% to 43% and 1% to 30%, respectively) intracranial hemorrhage and vessel perforation or dissection (11 studies; 0% to 7% and 0% to 5%, respectively), also varied by device. Predictors of harm included older age, history of stroke, and higher baseline stroke severity scores, whereas successful recanalization was the sole predictor of good outcomes. Limitations: Most available data are from single-group, noncomparative studies. In addition, the patient population most likely to benefit from these devices is undetermined. Conclusion: Currently available neurothrombectomy devices offer intriguing treatment options in patients with acute ischemic stroke. Future trials should use a randomized design, with adequate power to show equivalency or noninferiority between competing strategies or devices, and strive to identify populations that are most likely to benefit from use of neurothrombectomy devices.
引用
收藏
页码:243 / 252
页数:10
相关论文
共 62 条
  • [1] Arterial reocclusion in stroke patients treated with intravenous tissue plasminogen activator
    Alexandrov, AV
    Grotta, JC
    [J]. NEUROLOGY, 2002, 59 (06) : 862 - 867
  • [2] [Anonymous], 510(k) Premarket Notification
  • [3] BAKER WL, 2010, NEUROTHROMBECTOMY DE
  • [4] Approval of the MERCI Clot Retriever - A critical view
    Becker, KJ
    Brott, TG
    [J]. STROKE, 2005, 36 (02) : 400 - 403
  • [5] Mechanical thrombolysis in acute ischemic stroke with endovascular photoacoustic recanalization
    Berlis, A
    Lutsep, H
    Barnwell, S
    Norbash, A
    Wechsler, L
    Jungreis, CA
    Woolfenden, A
    Redekop, G
    Hartmann, M
    Schumacher, M
    [J]. STROKE, 2004, 35 (05) : 1112 - 1116
  • [6] The penumbra system: A mechanical device for the treatment of acute stroke due to thromboembolism
    Bose, A.
    Henkes, H.
    Alfke, K.
    Reith, W.
    Mayer, T. E.
    Berlis, A.
    Branca, V.
    Sit, S. Po
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2008, 29 (07) : 1409 - 1413
  • [7] Use of the New Solitaire™ AB Device for Mechanical Thrombectomy when Merci Clot Retriever Has Failed to Remove the Clot A Case Report
    Castano, C.
    Serena, J.
    Davalos, A.
    [J]. INTERVENTIONAL NEURORADIOLOGY, 2009, 15 (02): : 209 - 214
  • [8] Mechanical Thrombectomy With the Solitaire AB Device in Large Artery Occlusions of the Anterior Circulation A Pilot Study
    Castano, Carlos
    Dorado, Laura
    Guerrero, Cristina
    Millan, Monica
    Gomis, Meritxell
    Perez de la Ossa, Natalia
    Castellanos, Mar
    Rosa Garcia, M.
    Domenech, Sira
    Davalos, Antoni
    [J]. STROKE, 2010, 41 (08) : 1836 - 1840
  • [9] Validity and reliability of the Nintendo Wii Balance Board for assessment of standing balance
    Clark, Ross A.
    Bryant, Adam L.
    Pua, Yonghao
    McCrory, Paul
    Bennell, Kim
    Hunt, Michael
    [J]. GAIT & POSTURE, 2010, 31 (03) : 307 - 310
  • [10] Training Guidelines for Intra-arterial Catheter-directed Treatment of Acute Ischemic Stroke: A Statement from a Special Writing Group of the Society of Interventional Radiology
    Connors, J. J., III
    Sacks, David
    Black, Carl M.
    McIff, E. Bruce
    Stallmeyer, M. J. B.
    Cole, John W.
    Rowley, Howard A.
    Wojak, Joan C.
    Mericle, Robert A.
    Murphy, Kieran J.
    Cardella, John F.
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 20 (12) : 1507 - 1522