Enzyme coated glass pH-electrode: Its fabrication and applications in the determination of urea in blood samples

被引:22
作者
Sahney, R
Puri, BK
Anand, S
机构
[1] Indian Inst Technol, Ctr Biomed Engn, New Delhi 110016, India
[2] Indian Inst Technol, Dept Chem, New Delhi 110016, India
关键词
enzyme-coated electrode; urea sensor; sol-gel technique; urea determination; blood samples;
D O I
10.1016/j.aca.2005.03.069
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new enzyme coated electrode for the determination of urea in blood samples has been developed. It is based on the encapsulation of urease enzyme in the porous silicate matrix by the sol-gel technique on a glass electrode for the purpose of sensing urea in blood samples. Various parameters like the effect of pH, selection of a suitable buffer of appropriate concentration and interference of common substances in blood samples have been evaluated to optimize the conditions for the determination of urea. The electrode can be used for the determination of urea in the concentration range 0.03-30.0 mM in a solution. The detection limit of the present enzyme-coated electrode is found to be 52 mu g/ml of urea. The relative standard deviation for the electrode-to-electrode reproducibility is found to be 2.4 % for the determination of 0.1 mM of urea (six replicate electrodes). Sol-gel matrix containing immobilized enzyme was stable for about 25 days at similar to 4 degrees C with 80 % urease activity. Urea content in various clinical blood samples has been estimated using this electrode and the results are found to be in good agreement with the standard clinical methods as reported in the literature. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:157 / 161
页数:5
相关论文
共 20 条
[1]  
[Anonymous], QUANTITATIVE ANAL
[2]  
ANVIR D, 1994, CHEM MATER, V6, P1605
[3]   An amperometric urea biosensor based on a polyaniline-perfluorosulfonated ionomer composite electrode [J].
Cho, WJ ;
Huang, HJ .
ANALYTICAL CHEMISTRY, 1998, 70 (18) :3946-3951
[4]  
Das N, 1997, BIOTECHNOL BIOENG, V54, P329, DOI 10.1002/(SICI)1097-0290(19970520)54:4<329::AID-BIT5>3.0.CO
[5]  
2-M
[6]   SOL-GEL ENCAPSULATION METHODS FOR BIOSENSORS [J].
DAVE, BC ;
DUNN, B ;
VALENTINE, JS ;
ZINK, JI .
ANALYTICAL CHEMISTRY, 1994, 66 (22) :A1120-A1127
[7]  
ECKFELDT J, 1982, CLIN CHEM, V28, P1500
[8]   AMPEROMETRIC DETERMINATION OF UREA USING AN NADH-DEPENDENT COUPLED ENZYME [J].
GUILBAULT, GG ;
SEO, ML .
TALANTA, 1994, 41 (06) :1029-1033
[9]   THE SOL-GEL PROCESS [J].
HENCH, LL ;
WEST, JK .
CHEMICAL REVIEWS, 1990, 90 (01) :33-72
[10]   BIOSENSOR BASED ON DIRECT-DETECTION OF MEMBRANE-POTENTIAL INDUCED BY IMMOBILIZED HYDROLYTIC ENZYMES [J].
IVNITSKII, DM ;
RISHPON, J .
ANALYTICA CHIMICA ACTA, 1993, 282 (03) :517-525