Acute in vivo testing of an intravascular respiratory support catheter

被引:19
作者
Golob, JF
Federspiel, WJ
Merrill, TL
Frankowski, BJ
Litwak, K
Russian, H
Hattler, BG
机构
[1] Univ Pittsburgh, McGowan Ctr Artificial Organ Dev, Dept Surg, Pittsburgh, PA USA
[2] Univ Pittsburgh, Dept Bioengn, Pittsburgh, PA USA
[3] Univ Pittsburgh, Dept Chem Engn, Pittsburgh, PA 15261 USA
关键词
D O I
10.1097/00002480-200109000-00003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Current treatment for acute respiratory failure (ARF) includes the use of mechanical ventilation and/or extracorporeal membrane oxygenation, both of which can exacerbate lung injury. Intravenous respiratory support, using hollow fiber membranes placed in the vena cava, represents an attractive potential treatment for ARF, which could help reduce or eliminate ventilator induced trauma and/or other problems. Our group has been developing a respiratory support catheter (the Hattler catheter [HC]) that consists of a constrained hollow fiber bundle with a centrally located balloon. The balloon can be pulsated rapidly to increase blood flow across the fibers and decrease diffusional transfer resistance there, thus increasing gas exchange. The purpose of this study was to evaluate the HC in acute animal implants and to compare performance with that achieved in previous ex vivo studies. The HC was implanted into four calves by means of the external jugular vein and placed in the superior and inferior vena cava spanning the right atrium. Gas exchange, hemodynamics, and hematologic parameters were assessed over a range of balloon pulsation rates from 30 to 300 beats/minute. A <10% reduction in cardiac output was associated with catheter insertion and operation. The maximum CO, exchange rate occurred at the highest pulsation rate and averaged 56 +/- 3 ml/min, or 327 +/- 15 ml/min per m(2) when averaged to catheter membrane area, a level comparable to that achieved in the previous ex vivo studies. Balloon pulsation did not produce significant levels of hemolysis, as plasma-free hemoglobin remained below 10-15 mg/dl.
引用
收藏
页码:432 / 437
页数:6
相关论文
共 18 条
[1]  
Bray GA, 1999, DIABETES CARE, V22, P623
[2]   MAJOR FINDINGS FROM THE CLINICAL-TRIALS OF THE INTRAVASCULAR OXYGENATOR [J].
CONRAD, SA ;
BAGLEY, A ;
BAGLEY, B ;
SCHAAP, RN ;
GENTILELLO, L ;
TUCHSCHMIDT, J ;
STAIN, S ;
WEAVER, F ;
SNIDER, M ;
ZWISCHENBERGER, J ;
SHAPIRO, B ;
BARTLETT, R ;
MURRAY, M ;
HARRISON, B ;
SAMUELSON, W ;
LOCICERO, J ;
KIRBY, T ;
NAUNHEIM, K ;
BENNETT, D ;
FRIEDEL, N ;
BRUNET, F ;
BINDER, S ;
WOOD, A ;
WEBSTER, N ;
PAES, L ;
AREN, C ;
BENITO, S ;
VONSEGESSER, L ;
FIGUERA, D ;
JANSEN, EWL ;
HAVERICH, A ;
EVANS, T ;
ROSSAINT, R ;
PEPPERMAN, M ;
TANZEEM, A ;
KREYMANN, G ;
BETHUNE, D ;
MENTEC, H ;
ROUPIE, E ;
IVERSEN, S ;
HEDSTRAND, U .
ARTIFICIAL ORGANS, 1994, 18 (11) :846-863
[3]  
Cox C S Jr, 1991, ASAIO Trans, V37, pM411
[4]   Ex vivo testing of the intravenous membrane oxygenator [J].
Federspiel, WJ ;
Golob, JF ;
Merrill, TL ;
Lund, LW ;
Bultman, JA ;
Frankowski, BJ ;
Watach, M ;
Litwak, K ;
Hattler, BG .
ASAIO JOURNAL, 2000, 46 (03) :261-267
[5]  
Federspiel WJ, 1997, ASAIO J, V43, pM725
[6]  
Guyton A C., TXB MED PHYSL
[7]   DEVELOPMENT OF AN INTRAVENOUS MEMBRANE-OXYGENATOR - ENHANCED INTRAVENOUS GAS-EXCHANGE THROUGH CONVECTIVE MIXING OF BLOOD AROUND HOLLOW-FIBER MEMBRANES [J].
HATTLER, BG ;
REEDER, GD ;
SAWZIK, PJ ;
LUND, LW ;
WALTERS, FR ;
SHAH, AS ;
RAWLEIGH, J ;
GOODE, JS ;
KLAIN, M ;
BOROVETZ, HS .
ARTIFICIAL ORGANS, 1994, 18 (11) :806-812
[8]   A mathematical model of gas exchange in an intravenous membrane oxygenator [J].
Hewitt, TJ ;
Hattler, BG ;
Federspiel, WJ .
ANNALS OF BIOMEDICAL ENGINEERING, 1998, 26 (01) :166-178
[9]  
MORTENSEN JD, 1987, T AM SOC ART INT ORG, V33, P570
[10]  
Nguyen T T, 1993, ASAIO J, V39, pM719