Improved operational stability of biosensors based on enzyme-polyelectrolyte complex adsorbed into a porous carbon electrode

被引:49
作者
Gavalas, VG
Chaniotakis, NA [1 ]
Gibson, TD
机构
[1] Univ Crete, Dept Chem, Analyt Chem Lab, Iraklion 71409, Crete, Greece
[2] Univ Leeds, Dept Biochem & Mol Biol, Enzyme Biotechnol Grp, Leeds LS2 9JT, W Yorkshire, England
关键词
porous carbon electrode; polyelectrolyte; operational stability; glucose oxidase; horseradish peroxidase;
D O I
10.1016/S0956-5663(98)00066-9
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A novel porous active carbon is utilized in order to adsorb the diethylaminoethyl-dextran (DEAE-dextran)-enzyme stabilized complexes, for the construction of highly stable biosensors. The interaction of DEAE-dextran with the examined enzymes increases dramatically the operational stabilization of the sensors, without adverse effects on the enzyme activity. At the same time, the porous active carbon allows for high enzyme loading, good electrical contact and low resistance throughout the sensing element. Glucose oxidase and horseradish peroxidase are used as model enzymes in this study to construct biosensors, with very good reproducibility (less than 5% RSD). As a result, the glucose sensor exhibits very long operational stability (over a period of 5 months), while the hydrogen peroxide sensor retains its initial activity after several weeks. (C) 1998 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:1205 / 1211
页数:7
相关论文
共 12 条
  • [1] APPLETON A, 1997, SENSOR ACTUAT B-CHEM, V43, P65
  • [2] Cass A. E. G., 1990, BIOSENSORS PRACTICAL
  • [3] CSOREGI E, 1994, THESIS LUND U
  • [4] BIOCHEMICAL SENSORS - THE STATE-OF-THE-ART
    DELAGUARDIA, M
    [J]. MIKROCHIMICA ACTA, 1995, 120 (1-4) : 243 - 255
  • [5] USE OF POLYMER-FILMS IN AMPEROMETRIC BIOSENSORS
    EMR, SA
    YACYNYCH, AM
    [J]. ELECTROANALYSIS, 1995, 7 (10) : 913 - 923
  • [6] CARBON-PASTE ELECTRODES MODIFIED WITH ENZYMES, TISSUES, AND CELLS
    GORTON, L
    [J]. ELECTROANALYSIS, 1995, 7 (01) : 23 - 45
  • [7] AMPEROMETRIC BIOSENSORS BASED ON AN APPARENT DIRECT ELECTRON-TRANSFER BETWEEN ELECTRODES AND IMMOBILIZED PEROXIDASES - PLENARY LECTURE
    GORTON, L
    JONSSONPETTERSSON, G
    CSOREGI, E
    JOHANSSON, K
    DOMINGUEZ, E
    MARKOVARGA, G
    [J]. ANALYST, 1992, 117 (08) : 1235 - 1241
  • [8] Design of enzyme electrodes for extended use and storage life
    Khan, GF
    Wernet, W
    [J]. ANALYTICAL CHEMISTRY, 1997, 69 (14) : 2682 - 2687
  • [9] EFFECTS OF DIFFERENT ADDITIVES ON A TYROSINASE BASED CARBON-PASTE ELECTRODE
    LUTZ, M
    BURESTEDT, E
    EMNEUS, J
    LIDEN, H
    GOBHADI, S
    GORTON, L
    MARKOVARGA, G
    [J]. ANALYTICA CHIMICA ACTA, 1995, 305 (1-3) : 8 - 17
  • [10] Solid-contact ion-selective electrode with stable internal electrode
    Vamvakaki, M
    Chaniotakis, NA
    [J]. ANALYTICA CHIMICA ACTA, 1996, 320 (01) : 53 - 61