An interlaboratory study to test instrument performance of hydrogen dual-inlet isotope-ratio mass spectrometers

被引:23
作者
Brand, WA
Coplen, TB [1 ]
机构
[1] US Geol Survey, Natl Ctr 431, Reston, VA 20192 USA
[2] Max Planck Inst Biogeochem, D-07701 Jena, Germany
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 2001年 / 370卷 / 04期
关键词
D O I
10.1007/s002160100814
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An interlaboratory comparison of forty isotope-ratio mass spectrometers of different ages from several vendors has been performed to test H-2/H-1 performance with hydrogen gases of three different isotopic compositions. The isotope-ratio results (unsufficiently corrected for H-3(+) contribution to the m/z = 3 collector, uncorrected for valve leakage in the change-over valves, etc.) expressed relative to one of these three gases covered a wide range of values: -630 parts per thousand to -79o parts per thousand for the second gas and -368 parts per thousand to -462 parts per thousand for the third gas. After normalizing the isotopic abundances of these test gases (linearly adjusting the delta values so that the gases with the lowest and highest H-2 content were identical for all laboratories), the standard deviation of the 40 measurements of the intermediate gas was a remarkably low 0.85 parts per thousand. It is concluded that the use of scaling factors is mandatory for providing accurate internationally comparable isotope-abundance values. Linear scaling for the isotope-ratio scales of gaseous hydrogen mass spectrometers is completely adequate.
引用
收藏
页码:358 / 362
页数:5
相关论文
共 14 条
[1]   CONVERSION OF HYDROGENIC MATERIALS TO HYDROGEN FOR ISOTOPIC ANALYSIS [J].
BIGELEISEN, J ;
PERLMAN, ML ;
PROSSER, HC .
ANALYTICAL CHEMISTRY, 1952, 24 (08) :1356-1357
[2]   REDUCTION OF WATER WITH ZINC FOR HYDROGEN ISOTOPE ANALYSIS [J].
COLEMAN, ML ;
SHEPHERD, TJ ;
DURHAM, JJ ;
ROUSE, JE ;
MOORE, GR .
ANALYTICAL CHEMISTRY, 1982, 54 (06) :993-995
[3]   IMPROVEMENTS IN THE GASEOUS-HYDROGEN WATER EQUILIBRATION TECHNIQUE FOR HYDROGEN ISOTOPE RATIO ANALYSIS [J].
COPLEN, TB ;
WILDMAN, JD ;
CHEN, J .
ANALYTICAL CHEMISTRY, 1991, 63 (09) :910-912
[4]   NORMALIZATION OF OXYGEN AND HYDROGEN ISOTOPE DATA [J].
COPLEN, TB .
CHEMICAL GEOLOGY, 1988, 72 (04) :293-297
[5]   DEUTERIUM CONTENT OF NATURAL WATERS AND OTHER SUBSTANCES [J].
FRIEDMAN, I .
GEOCHIMICA ET COSMOCHIMICA ACTA, 1953, 4 (1-2) :89-103
[6]   Sample preparation device for quantitative hydrogen isotope analysis using chromium metal [J].
Gehre, M ;
Hoefling, R ;
Kowski, P ;
Strauch, G .
ANALYTICAL CHEMISTRY, 1996, 68 (24) :4414-4417
[7]   STANDARDS FOR STABLE ISOTOPE MEASUREMENTS IN NATURAL COMPOUNDS [J].
GONFIANTINI, R .
NATURE, 1978, 271 (5645) :534-536
[8]  
HORITA J, 1989, APPL RADIAT ISOTOPES, V40, P801, DOI 10.1016/0883-2889(89)90100-7
[9]   HYDROGEN ISOTOPE ANALYSIS OF NATURAL-WATERS USING AN H-2 WATER EQUILIBRATION METHOD - A SPECIAL IMPLICATION TO BRINES [J].
HORITA, J .
CHEMICAL GEOLOGY, 1988, 72 (01) :89-94
[10]  
KIRSHENBAUM I, 1951, PHYSICAL PROPERTIES, P438