Effects of mechanical fatigue on the bending properties of the porcine bioprosthetic heart valve

被引:70
作者
Gloeckner, DC [1 ]
Billiar, KL [1 ]
Sacks, MS [1 ]
机构
[1] Univ Miami, Dept Biomed Engn, Tissue Mech Lab, Miami, FL 33152 USA
关键词
D O I
10.1097/00002480-199901000-00014
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The mechanisms underlying the failure of porcine bioprosthetic aortic heart valves are not well understood. One possible explanation is that delaminations of the layered leaflet structure occur through flexion, leading to calcification and further delaminations, and finally resulting in valve failure. We investigated the changes in flexural rigidity of the belly of aortic valve cusps subjected to accelerated durability testing. We used three-point bending wherein a load was applied to the center of each specimen by a thin stainless steel bar calibrated to a known load-displacement relationship. Ten circumferential and 15 radial specimens from valves fatigued to 0, 50, 100, and 200 million cycles were flexed both with and against the curvature of the cusp. Linear beam theory was applied as a means to compare the relative bending stiffness between groups. Although specimens aligned to the circumferential direction were stiffer when bent against the cuspal curvature, the radial oriented specimens exhibited no bending directional dependence. Both the radial and circumferential specimens experienced a significant decrease in the bending stiffness with an increased number of accelerated test cycles. Overall, our results suggest that it is the fibrosa that experiences the greatest loss of stiffness with mechanically induced fatigue damage.
引用
收藏
页码:59 / 63
页数:5
相关论文
共 13 条
[1]  
BILLIAR K, 1998, IMECEX 98 AN
[2]  
Haziza F, 1996, J HEART VALVE DIS, V5, P35
[3]  
POMAR JL, 1984, ANN THORAC SURG, V37, P83
[4]  
*PROSTH DEV BRANC, 1993, RESP NEUR DEV REPL H
[5]  
Schoen F J, 1992, Cardiovasc Pathol, V1, P29, DOI 10.1016/1054-8807(92)90006-A
[6]  
Schoen FJ, 1997, J HEART VALVE DIS, V6, P1
[7]  
Thubrikar M, 1990, AORTIC VALVE, P221
[8]   STRESS-ANALYSIS OF PORCINE BIOPROSTHETIC HEART-VALVES INVIVO [J].
THUBRIKAR, MJ ;
SKINNER, JR ;
EPPINK, RT ;
NOLAN, SP .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1982, 16 (06) :811-826
[9]  
THUBRIKAR MJ, 1983, J THORAC CARDIOV SUR, V86, P115
[10]   A MULTIPURPOSE TISSUE BENDING MACHINE [J].
VESELY, I ;
BOUGHNER, DR .
JOURNAL OF BIOMECHANICS, 1985, 18 (07) :511-513