The AAPM/RSNA physics tutorial for residents - Digital fluoroscopy

被引:22
作者
Pooley, RA [1 ]
McKinney, JM [1 ]
Miller, DA [1 ]
机构
[1] Mayo Clin, Dept Radiol, Jacksonville, FL 32224 USA
关键词
digital subtraction angiography; fluoroscopy; images; processing; physics; radiography; digital;
D O I
10.1148/radiographics.21.2.g01mr20521
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
A digital fluoroscopy system is most commonly configured as a conventional fluoroscopy system (tube, table, image intensifier, video system) in which the analog video signal is converted to and stored as digital data. Other methods of acquiring the digital data (eg, digital or charge-coupled device video and flat-panel detectors) will become more prevalent in the future. Fundamental concepts related to digital imaging in general include binary numbers, pixels, and gray levels. Digital image data allow the convenient use of several image processing techniques including last image hold, gray-scale processing, temporal frame averaging, and edge enhancement. Real-time subtraction of digital fluoroscopic images after injection of contrast material has led to widespread use of digital subtraction angiography (DSA). Additional image processing techniques used with DSA include road mapping, image fade, mask pixel shift, frame summation, and vessel size measurement. Peripheral angiography performed with an automatic moving table allows imaging of the peripheral vasculature with a single contrast material injection.
引用
收藏
页码:521 / 534
页数:14
相关论文
共 5 条
[1]
HOFFMANN KR, 1995, SYLLABUS CATEGORICAL, P143
[2]
KRUGER RA, 1979, RADIOLOGY, V130, P49, DOI 10.1148/130.1.49
[3]
SEIBERT JA, 1995, SYLLABUS CATEGORICAL, P121
[4]
SPRAWLS P, 1997, EXPANDING ROLE MED P, P18
[5]
VERHOEVEN L, 1994, SPECIFICATION ACCEPT, P651