Analysis and generalized correction of the effect of spatial gradient field distortions in diffusion-weighted imaging

被引:128
作者
Bammer, R
Markl, M
Barnett, A
Acar, B
Alley, MT
Pelc, NJ
Glover, GH
Moseley, ME
机构
[1] Stanford Univ, Dept Radiol, Richard Lucas MRSI Ctr, Radiol Sci Lab, Stanford, CA 94305 USA
[2] NIMH, Clin Brain Disorders Branch, CBDB, Bethesda, MD 20892 USA
关键词
diffusion-weighted imaging; diffusion tensor imaging; magnetic field gradients; data correction; MRI; fiber tracking;
D O I
10.1002/mrm.10545
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Nonuniformities of magnetic field gradients can cause serious artifacts in diffusion imaging. While it is well known that nonlinearities of the imaging gradients lead to image warping, those imperfections can also cause spatially dependent errors in the direction and magnitude of the diffusion encoding. This study shows that the potential errors in diffusion imaging are considerable. Further, we show that retrospective corrections can be applied to reduce these errors. A general mathematical framework was formulated to characterize the contribution of gradient nonuniformities to diffusion experiments. The gradient field was approximated using spherical harmonic expansion, and this approximation was employed (after geometric distortions were eliminated) to predict and correct the errors in diffusion encoding. Before the corrections were made, the experiments clearly revealed marked deviations of the calculated diffusivity for fields of view (FOVs) generally used in diffusion experiments. These deviations were most significant farther away from the magnet's isocenter. For an FOV of 25 cm, the resultant errors in absolute diffusivity ranged from approximately -10% to +20%. Within the same FOV, the diffusion-encoding direction and the orientation of the calculated eigenvectors can be significantly altered if the perturbations by the gradient nonuniformities are not considered. With the proposed correction scheme, most of the errors introduced by gradient nonuniformities can be removed. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:560 / 569
页数:10
相关论文
共 21 条
[1]   Spatial transformations of diffusion tensor magnetic resonance images [J].
Alexander, DC ;
Pierpaoli, C ;
Basser, PJ ;
Gee, JC .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2001, 20 (11) :1131-1139
[2]   Diffusion imaging in multiple sclerosis [J].
Bammer, R ;
Fazekas, F .
NEUROIMAGING CLINICS OF NORTH AMERICA, 2002, 12 (01) :71-+
[3]  
Bammer R, 2002, PROCEED INT SOC MAGN, P1172
[4]  
BAMMER R, 2001, P 9 ANN M ISMRM GLAS, P508
[5]  
Basser PJ, 1996, J MAGN RESON SER B, V111, P209, DOI [10.1006/jmrb.1996.0086, 10.1016/j.jmr.2011.09.022]
[6]   Concomitant gradient terms in phase contrast MR: Analysis and correction [J].
Bernstein, MA ;
Zhou, XHJ ;
Polzin, JA ;
King, KF ;
Ganin, A ;
Pelc, NJ ;
Glover, GH .
MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (02) :300-308
[7]  
Bernstein MA, 2000, US Patent, Patent No. [6,163,152, 6163152]
[8]   Diffusion MRI: Precision, accuracy and flow effects [J].
Conturo, TE ;
McKinstry, RC ;
Aronovitz, JA ;
Neil, JJ .
NMR IN BIOMEDICINE, 1995, 8 (7-8) :307-332
[9]   Anisotropy in high angular resolution diffusion-weighted MRI [J].
Frank, LR .
MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (06) :935-939
[10]  
Glover G., 1986, US Patent, Patent No. [4,591,789., 4591789, 4,591,789]