Use of overpressure to assess the role of bubbles in focused ultrasound lesion shape in vitro

被引:131
作者
Bailey, MR
Couret, LN
Sapozhnikov, OA
Khokhlova, VA
Ter Haar, G
Vaezy, S
Shi, XG
Martin, R
Crum, LA
机构
[1] Univ Washington, Appl Phys Lab, Seattle, WA 98105 USA
[2] Univ Washington, Dept Bioengn, Seattle, WA 98195 USA
[3] Univ Washington, Dept Anesthesiol, Seattle, WA 98195 USA
[4] Moscow State Univ, Fac Phys, Dept Acoust, Moscow, Russia
[5] Royal Marsden Hosp, Inst Canc Res, Joint Dept Phys, Surrey, England
关键词
lesion; ultrasound; cavitation; overpressure; HIFU;
D O I
10.1016/S0301-5629(01)00342-8
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Overpressure-elevated hydrostatic pressure-was used to assess the role of gas or vapor bubbles in distorting the shape and position of a high-intensity focused ultrasound (HIFU) lesion in tissue, The shift from a cigar-shaped lesion to a tadpole-shaped lesion can mean that the wrong area is treated. Overpressure minimizes hubbies and bubble activity by dissolving gas bubbles, restricting bubble oscillation and raising the boiling temperature, Therefore, comparison with and without overpressure is a tool to assess the role of bubbles. Dissolution rates, bubble dynamics and boiling temperatures were determined as functions of pressure. Experiments were made first in a low-overpressure chamber (0.7 MPa maximum) that permitted imaging by B-mode ultrasound (US), Pieces of excised beef liver (8 cm thick) were treated in the chamber with 3.5 MHz for 1 to 7 s (50% duty cycle). In situ intensities (I-SP) were 600 to 3000 W/cm(2). B-mode US imaging detected a hyperechoic region at the HIFU treatment site. The dissipation of this hyperechoic region following HIFU cessation corresponded well with calculated bubble dissolution rates; thus, suggesting that bubbles were present, Lesion shape was then tested in a high-pressure chamber. Intensities were 1300 and 1750 W/cm(2) (+/- 20%) at 1 MHz for 30 s, Hydrostatic pressures were 0.1 or 5.6 MPa, At 1300 W/cm(2), lesions were cigar-shaped, and no difference was observed between lesions formed with or without overpressure, At 1750 W/cm(2), lesions formed with no overpressure were tadpole-shaped, but lesions formed with high overpressure (5.6 MPa) remained cigar-shaped. Data support the hypothesis that bubbles contribute to the lesion distortion. (E-mail: bailey@apl.washington.edu) (C) 2001 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:695 / 708
页数:14
相关论文
共 51 条
[1]  
Akulichev V. A., 1971, HIGH INTENSITY ULTRA, P239
[2]  
AKULICHEV VA, 1966, SOV PHYS ACOUST+, V12, P144
[3]  
[Anonymous], J APPL PHYSL
[5]   THRESHOLDS FOR CAVITATION PRODUCED IN WATER BY PULSED ULTRASOUND [J].
ATCHLEY, AA ;
FRIZZELL, LA ;
APFEL, RE ;
HOLLAND, CK ;
MADANSHETTY, S ;
ROY, RA .
ULTRASONICS, 1988, 26 (05) :280-285
[6]  
ATCHLEY AA, 1984, ACOUSTIC CAVITATION, P120
[7]  
BAILEY MR, 1999, JOINT M BERL 99 137
[8]  
BRONSKAYA LM, 1968, SOV PHYS ACOUST+, V13, P374
[9]   Modeling of high-intensity focused ultrasound-induced lesions in the presence of cavitation bubbles [J].
Chavrier, F ;
Chapelon, JY ;
Gelet, A ;
Cathignol, D .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 108 (01) :432-440
[10]   A THEORETICAL-STUDY OF CAVITATION GENERATED BY AN EXTRACORPOREAL SHOCK-WAVE LITHOTRIPTER [J].
CHURCH, CC .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1989, 86 (01) :215-227