Effects of stent porosity on hemodynamics in a sidewall aneurysm model

被引:112
作者
Liou, Tong-Miin [1 ]
Li, Yi-Chen [1 ]
机构
[1] Natl Tsing Hua Univ, Hsinchu 30013, Taiwan
关键词
stented lateral aneurysms; porosity; wall shear stress; pressure; flow rate waveform;
D O I
10.1016/j.jbiomech.2008.01.025
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Computation and experiment have been complementarily performed to study the fluid flow inside a stented lateral aneurysm anchored on the straight parent vessel. The implicit solver was based on the time-dependent incompressible Navier-Stokes equations of laminar flow. Solutions were generated by a cell-center finite-volume method that used second-order upwind and second-order center flux difference splitting for the convection and diffusion term, respectively. The second-order Crank-Nicolson method was used in the time integration term. Experimental techniques used were flow visualization (FV) and particle tracking velocimetry (PTV). Experimentally, the straight afferent vessel had an inner diameter 10mm. The diameters of the aneurysmal orifice, neck, and fundus were 14, 10, and 15 mm, respectively, and the distance between the orifice and dome measured 20 mm. A 30 mm long helix-shaped stent was tested. Four stent porosities of 100%, 70%, 50%, and 25% were examined. Volume-flow rate waveform of a cerebral artery was considered with a maximum Reynolds number of 250 and Womersley number of 3.9. Results are presented in terms of the pulsatile main and secondary flow velocity vector fields, the volume inflow rates into the aneurysm, and the wall shear stress (WSS) and wall pressure at the aneurysm dome. Some comparisons of computed results with the present FV and PTV results and with the data available from the literature are also made. The maximum flow velocity inside the aneurysm ostium and the WSS in the dome region at the peak flow can, respectively, be suppressed to less than 5% of the parent vessel bulk velocity (or 20% of the unstented case) and 8% of the unstented case if the stent porosity is smaller than.40% (about the porosity of the two-layer stents). In general, the three-layer stents seem not as effective as the two-layer stents in reducing the magnitude of aneurysm inflow rate and WSS. (c) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1174 / 1183
页数:10
相关论文
共 23 条
[1]   Modeling of flow in a straight stented and nonstented side wall aneurysm model [J].
Aenis, M ;
Stancampiano, AP ;
Wakhloo, AK ;
Lieber, BB .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1997, 119 (02) :206-212
[2]   Life-threatening bleeding from a vertebral artery pseudoaneurysm after anterior cervical spine approach: endovascular repair by a triple stent-in-stent method. Case report [J].
Alzamora, MG ;
Rosahl, SK ;
Lehmberg, J ;
Klisch, J .
NEURORADIOLOGY, 2005, 47 (04) :282-286
[3]  
BANDO K, 2003, COMP FLUID DYNAM J, V11, P527
[4]  
Benndorf G, 2001, AM J NEURORADIOL, V22, P1844
[5]   FLOW STIMULATES ENDOTHELIAL-CELLS TO RELEASE A NITROVASODILATOR THAT IS POTENTIATED BY REDUCED THIOL [J].
COOKE, JP ;
STAMLER, J ;
ANDON, N ;
DAVIES, PF ;
MCKINLEY, G ;
LOSCALZO, J .
AMERICAN JOURNAL OF PHYSIOLOGY, 1990, 259 (03) :H804-H812
[7]   TRANSLUMINAL TREATMENT OF ARTERIOSCLEROTIC OBSTRUCTION - DESCRIPTION OF NEW TECHNIC + PRELIMINARY REPORT OF ITS APPLICATION [J].
DOTTER, CT ;
JUDKINS, MP .
CIRCULATION, 1964, 30 (05) :654-&
[8]  
GEREMIA G, 1994, AM J NEURORADIOL, V15, P1223
[9]   IN-VITRO STUDY OF HEMODYNAMICS IN A GIANT SACCULAR ANEURYSM MODEL - INFLUENCE OF NOW DYNAMICS IN THE PARENT VESSEL AND EFFECTS OF COIL EMBOLIZATION [J].
GOBIN, YP ;
COUNORD, JL ;
FLAUD, P ;
DUFFAUX, J .
NEURORADIOLOGY, 1994, 36 (07) :530-536
[10]   Alteration of hemodynamics in aneurysm models by stenting: Influence of stent porosity [J].
Lieber, BB ;
Stancampiano, AP ;
Wakhloo, AK .
ANNALS OF BIOMEDICAL ENGINEERING, 1997, 25 (03) :460-469