Sensitivity deviation:: instrumental linearity errors that influence concentration analyses and kinetic evaluation of biomolecular interactions

被引:12
作者
Johansen, K [1 ]
Lundström, I [1 ]
Liedberg, B [1 ]
机构
[1] Linkoping Univ, Dept Phys & Measurement Technol, Appl Phys Lab, S-58183 Linkoping, Sweden
基金
瑞典研究理事会;
关键词
linearity error; sensitivity deviation; surface plasmon resonance; kinetic evaluation; concentration analysis;
D O I
10.1016/S0956-5663(00)00109-3
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Many scientific instruments utilise multiple element detectors, e.g. CCD's or photodiode arrays, to monitor the change in a position of an optical pattern. For example, instruments for affinity biosensing based on surface plasmon resonance (SPR) or resonant mirror are equipped with such detectors. An important and desired property of these bioanalytical instruments is that the calculation of the movement or change in shape follows the true change. This is often not the case and it may lead to linearity errors, and to sensitivity errors. The sensitivity is normally defined as the slope of the calibration curve. A new parameter is introduced to account for the linearity errors, the sensitivity deviation, defined as the deviation from the undistorted slope of the calibration curve. The linearity error and the sensitivity deviation are intimately related and the sensitivity deviation may lead to misinterpretation of kinetic data, mass transport limitations and concentration analyses. Because the linearity errors are small (e.g. 10 pg/mm(2) of biomolecules on the sensor surface) with regard to the dynamic range (e.g. 30000 pg/mm(2)), they can be difficult to discover. However, the linearity errors are often not negligible with regard to a typical response (e.g. 0-100 pg/mm(2)), and may therefore cause serious problems. A method for detecting linearity errors is outlined. Further on, this paper demonstrates how integral linearity errors of less than 1% can result in a sensitivity deviation of 10%, a value that in our opinion cannot be ignored in biospecific interaction analysis (BIA). It should also be stressed out that this phenomenon also occurs in other instruments using array detectors. (C) 2000 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:503 / 509
页数:7
相关论文
共 19 条
[1]  
BIACORE AB, 1997, BIA EVALUATION VERSI
[2]  
DOEBELIN EO, 1975, MEASUREMENT SYSTEMS
[3]   BIOSPECIFIC INTERACTION ANALYSIS USING SURFACE-PLASMON RESONANCE DETECTION APPLIED TO KINETIC, BINDING-SITE AND CONCENTRATION ANALYSIS [J].
FAGERSTAM, LG ;
FROSTELLKARLSSON, A ;
KARLSSON, R ;
PERSSON, B ;
RONNBERG, I .
JOURNAL OF CHROMATOGRAPHY, 1992, 597 (1-2) :397-410
[4]   ANTIGEN-ANTIBODY BINDING AND MASS-TRANSPORT BY CONVECTION AND DIFFUSION TO A SURFACE - A 2-DIMENSIONAL COMPUTER-MODEL OF BINDING AND DISSOCIATION KINETICS [J].
GLASER, RW .
ANALYTICAL BIOCHEMISTRY, 1993, 213 (01) :152-161
[5]   REAL-TIME BIOMOLECULAR INTERACTION ANALYSIS USING THE RESONANT MIRROR SENSOR [J].
GODDARD, NJ ;
POLLARDKNIGHT, D ;
MAULE, CH .
ANALYST, 1994, 119 (04) :583-588
[6]  
*ICH, 1996, ICH TOP Q 2 B ICH
[7]   KINETIC-ANALYSIS OF MONOCLONAL ANTIBODY-ANTIGEN INTERACTIONS WITH A NEW BIOSENSOR BASED ANALYTICAL SYSTEM [J].
KARLSSON, R ;
MICHAELSSON, A ;
MATTSSON, L .
JOURNAL OF IMMUNOLOGICAL METHODS, 1991, 145 (1-2) :229-240
[8]   Experimental design for kinetic analysis of protein-protein interactions with surface plasmon resonance biosensors [J].
Karlsson, R ;
Falt, A .
JOURNAL OF IMMUNOLOGICAL METHODS, 1997, 200 (1-2) :121-133
[9]  
KARLSSON R, 1993, J IMMUNOL METHODS, V166, P74
[10]  
Karlsson Robert, 1994, Methods (Orlando), V6, P99, DOI 10.1006/meth.1994.1013