Direct measurement of Cu(II)aq in seawater at pH 8 with the jalpaite ion-selective electrode

被引:42
作者
Zirino, A
VanderWeele, DA
Belli, SL
DeMarco, R
Mackey, DJ
机构
[1] SPAWAR Syst Ctr, San Diego, CA 92152 USA
[2] Curtin Univ Technol, Sch Appl Chem, Perth, WA 6001, Australia
[3] CSIRO, Div Oceanog, Hobart, Tas 7001, Australia
关键词
measurement; seawater; copper; ion-selective electrode;
D O I
10.1016/S0304-4203(97)00108-4
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The variability of measurement of the pCu of seawater at pH 8 was studied using up to 5 similar Orion 90-24, jalpaite-based electrodes. Large variability (ca. 1 pCu unit) among the values yielded by the individual electrodes was observed when the electrodes were stored in artificial seawater and exposed to light including sunlight This variability is in agreement with mechanisms proposed previously. The variability among the electrodes was reduced to about +/-0.2 pCu units when the electrodes were stored in sacrificial EN seawater solution and shielded from light. Eight measurements of San Diego Bay seawater made with an individual electrode yielded pCu = 12.23 +/-:0.06 (1 sigma) regardless of the order in which the electrode was calibrated with standard buffer solutions. This degree of precision makes it possible to follow the small, time-dependent changes in the pCu of seawater. The effects of electrode rotation and the adsorption or desorption of organic matter on the electrode are presented and discussed. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:173 / 184
页数:12
相关论文
共 22 条
[1]   BEHAVIOR AND CALIBRATION OF THE COPPER(II) ION-SELECTIVE ELECTRODE IN HIGH CHLORIDE MEDIA AND MARINE WATERS [J].
BELLI, SL ;
ZIRINO, A .
ANALYTICAL CHEMISTRY, 1993, 65 (19) :2583-2589
[2]   STUDY OF BEHAVIOR OF COPPER ION-SELECTIVE ELECTRODES AT SUBMICROMOLAR CONCENTRATION LEVELS [J].
BLAEDEL, WJ ;
DINWIDDI.DE .
ANALYTICAL CHEMISTRY, 1974, 46 (07) :873-877
[3]   COMPLEXATION BY NATURAL HETEROGENEOUS COMPOUNDS - SITE OCCUPATION DISTRIBUTION-FUNCTIONS, A NORMALIZED DESCRIPTION OF METAL COMPLEXATION [J].
BUFFLE, J ;
ALTMANN, RS ;
FILELLA, M ;
TESSIER, A .
GEOCHIMICA ET COSMOCHIMICA ACTA, 1990, 54 (06) :1535-1553
[4]  
Buffle J., 1988, Complexation Reactions in Aquatic Systems
[5]   Surface studies of the jalpaite-based copper(II) ion-selective electrode membrane in seawater [J].
DeMarco, R .
MARINE CHEMISTRY, 1996, 55 (3-4) :389-398
[6]   SURFACE STUDIES OF THE COPPER SILVER SULFIDE BASED ION-SELECTIVE ELECTRODE MEMBRANE [J].
DEMARCO, R ;
CATTRALL, RW ;
LIESEGANG, J ;
NYBERG, GL ;
HAMILTON, IC .
ANALYTICAL CHEMISTRY, 1992, 64 (06) :594-598
[7]   Response of the jalpaite membrane copper(II) ion-selective electrode in marine waters [J].
DeMarco, R ;
Mackey, DJ ;
Zirino, A .
ELECTROANALYSIS, 1997, 9 (04) :330-334
[8]  
DEMARCO R, 1994, ANAL CHEM, V66, P3203
[9]   Copper ligands isolated from estuarine water by immobilized metal affinity chromatography: Temporal variability and partial characterization [J].
Gordon, AS ;
Dyer, BJ ;
Kango, RA ;
Donat, JR .
MARINE CHEMISTRY, 1996, 53 (3-4) :163-172
[10]  
GULENS J, 1987, ION SEL ELECTRODE R, V9, P127