Failure analysis of broken pedicle screws on spinal instrumentation

被引:172
作者
Chen, CS
Chen, WJ
Cheng, CK [1 ]
Jao, SHE
Chueh, SC
Wang, CC
机构
[1] Natl Yang Ming Univ, Inst Rehabil Sci & Technol, Taipei 112, Taiwan
[2] Chang Gung Mem Hosp, Dept Orthoped, Taoyuan, Taiwan
[3] Natl Yang Ming Univ, Inst Biomed Engn, Orthopaed Biomech Lab, Taipei 112, Taiwan
[4] Natl Yang Ming Univ, Joint Prosthesis Technol Res Ctr, Taipei 112, Taiwan
[5] Ind Technol Res Inst, Hsinchu, Taiwan
关键词
pedicle screw; scanning electron microscope; finite element method; fatigue; bone union;
D O I
10.1016/j.medengphy.2004.12.007
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Revised spinal surgery is needed when there is a broken pedicle screw in the patient. This study investigated the pedicle screw breakage by conducting retrieval analyses of broken pedicle screws from 16 patients clinically and by performing stress analyses in the posterolateral fusion computationally using finite element (FE) models. Fracture surface of screws was studied by scanning electron microscope (SEM). The FE model of the posterolateral fusion with the screw showed that screws on the caudal side had larger axial stress than those on the cephalic side, supporting the clinical findings that 75% of the patients had the screw breakage on the caudal side. SEM fractography showed that all broken screws exhibited beach marks or striations on the fractured surface, indicating fatigue failure. Screws of patients with spinal fracture showed fatigue striations and final ductile fracture around the edge. Among the 16 patients who had broken pedicle screws 69% of them achieved bone union in the bone graft, showing that bone union in the bone graft did not warrant the prevention of screw breakage. (c) 2005 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:487 / 496
页数:10
相关论文
共 16 条
[1]
*ASM INT, 1986, MET HDB, P360
[2]
*ASM INT, 1986, MET HDB, P670
[3]
BUCKWALTER JA, 1991, FRACTURES, P181
[4]
A biomechanical comparison of posterolateral fusion and posterior fusion in the lumbar spine [J].
Chen, CS ;
Cheng, CK ;
Liu, CL ;
Simmons, ED .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2002, 15 (01) :53-63
[5]
Chen CS, 2001, MED ENG PHYS, V23, P483
[6]
CHEN CS, IN PRES J SPINAL DIS
[7]
Cotler JM, 1990, SPINAL FUSION SCI TE, P361
[8]
STATIC AND CYCLICAL BIOMECHANICAL ANALYSIS OF PEDICLE SCREW SPINAL CONSTRUCTS [J].
CUNNINGHAM, BW ;
SEFTER, JC ;
SHONO, Y ;
MCAFEE, PC .
SPINE, 1993, 18 (12) :1677-1688
[9]
TRANSPEDICULAR SCREW-ROD FIXATION OF THE LUMBAR SPINE - OPERATIVE TECHNIQUE AND OUTCOME IN 104 CASES [J].
DICKMAN, CA ;
FESSLER, RG ;
MACMILLAN, M ;
HAID, RW .
JOURNAL OF NEUROSURGERY, 1992, 77 (06) :860-870
[10]
EFFECTS OF RIGIDITY OF AN INTERNAL-FIXATION DEVICE - A COMPREHENSIVE BIOMECHANICAL INVESTIGATION [J].
GOEL, VK ;
LIM, TH ;
GWON, J ;
CHEN, JY ;
WINTERBOTTOM, JM ;
PARK, JB ;
WEINSTEIN, JN ;
AHN, JY .
SPINE, 1991, 16 (03) :S155-S161