Biosensor for the evaluation of biochemical oxygen demand using photocatalytic pretreatment

被引:16
作者
Chee, GJ [1 ]
Nomura, Y [1 ]
Ikebukuro, K [1 ]
Karube, I [1 ]
机构
[1] Univ Tokyo, Adv Sci & Technol Res Ctr, Meguro Ku, Tokyo 1538904, Japan
关键词
BOD sensor; biosensor; Pseudomonas putida; artificial wastewater; photocatalysis; TiO2; UV; pretreatment;
D O I
10.1016/S0925-4005(01)00883-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A highly sensitive biochemical oxygen demand (BOD) sensor using photocatalytic pretreatment was developed to evaluate low BOD levels in river waters. The photocatalytic oxidation was carried out in a commercial TiO2 as a semiconductor and a 6 W black-light tube as light sources. The photocatalytic oxidation was investigated about the effects of irradiation times, TiO2 concentrations, and pH on the sensor response. The optimal irradiation time was 4 min. TiO2 concentrations gave the optimum response to the BOD sensor at 1% (w/v). The sensor response was increased with increasing pH. The sensor responses obtained by photocatalysis to river samples were higher than those obtained without photocatalysis, A fairly good correlation between the values obtained by the photocatalytic pretreatment method and BOD5 was acquired. (C) 2001 Elsevier Science B.V. All rights reserved.