Performance evaluation of a recently developed water-based condensation particle counter

被引:37
作者
Biswas, S
Fine, PM
Geller, MD
Hering, SV
Sioutas, C
机构
[1] Univ So Calif, Dept Civil & Environm Engn, Los Angeles, CA 90089 USA
[2] Aerosol Dynam Inc, Berkeley, CA USA
关键词
D O I
10.1080/027868290953173
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
This study provides an intercomparison of the performance of a newly developed water-based condensation particle counter (WCPC) and a more widely used butanol-based CPC (TSI 3022A). Four test aerosols (ammonium nitrate, ammonium sulfate, adipic acid, and glutaric acid) were generated and tested in the laboratory before the instruments were deployed at four field locations (USC/downtown LA, I-710 Freeway, Pacific coast, and Los Angeles International Airport). Both instruments sampled the same incoming aerosol. Selected experiments utilized a differential mobility analyzer to select a particle size upstream of the CPCs. Evaluation of performance was based on the response of the instruments to varying particle composition, concentrations, and size. The results indicated good correlation between the two CPCs, with R-2 values ranging from 0.74-0.99. Good agreement was found between the two instruments for particle concentrations between 0 and 40,000 particles/cm(3), with W-CPC/TSI 3022A ratios between 0.8 and 1.2. Due to differences in the photometric mode calibration of these instruments, the ratio drops to 0.6-0.8 between 40,000-100,000 particles/cm(3). However, the ratio rises again for lab aerosols above 100,000 particles/cm(3) to 1.0-1.1. Results of this evaluation show that the W-CPC is a reliable particle-counting technology for particle concentrations encountered downstream of a DMA as well as in some ambient environments (< 40,000 particles/cm(3)).
引用
收藏
页码:419 / 427
页数:9
相关论文
共 43 条
[1]   Particle size dependent response of aerosol counters [J].
Ankilov, A ;
Baklanov, A ;
Colhoun, M ;
Enderle, KH ;
Gras, J ;
Julanov, Y ;
Kaller, D ;
Lindner, A ;
Lushnikov, AA ;
Mavliev, R ;
McGovern, F ;
O'Connor, TC ;
Podzimek, J ;
Preining, O ;
Reischl, GP ;
Rudolf, R ;
Sem, GJ ;
Szymanski, WW ;
Vrtala, AE ;
Wagner, PE ;
Winklmayr, W ;
Zagaynov, V .
ATMOSPHERIC RESEARCH, 2002, 62 (3-4) :209-237
[2]   RESPONSE CHARACTERISTICS FOR 4 DIFFERENT CONDENSATION NUCLEUS COUNTERS TO PARTICLES IN THE 3-50 NM DIAMETER RANGE [J].
BARTZ, H ;
FISSAN, H ;
HELSPER, C ;
KOUSAKA, Y ;
OKUYAMA, K ;
FUKUSHIMA, N ;
KEADY, PB ;
KERRIGAN, S ;
FRUIN, SA ;
MCMURRY, PH ;
PUI, DYH ;
STOLZENBURG, MR .
JOURNAL OF AEROSOL SCIENCE, 1985, 16 (05) :443-456
[3]  
Buzorius G, 2001, AEROSOL SCI TECH, V35, P577
[4]   Condensation of organic vapors on an externally mixed aerosol population [J].
Cruz, CN ;
Pandis, SN .
AEROSOL SCIENCE AND TECHNOLOGY, 1999, 31 (05) :392-407
[5]   AN ASSOCIATION BETWEEN AIR-POLLUTION AND MORTALITY IN 6 UNITED-STATES CITIES [J].
DOCKERY, DW ;
POPE, CA ;
XU, XP ;
SPENGLER, JD ;
WARE, JH ;
FAY, ME ;
FERRIS, BG ;
SPEIZER, FE .
NEW ENGLAND JOURNAL OF MEDICINE, 1993, 329 (24) :1753-1759
[6]   The particle detection efficiency of the TSI-3007 condensation particle counter [J].
Hämeri, K ;
Koponen, IK ;
Aalto, PP ;
Kulmala, M .
JOURNAL OF AEROSOL SCIENCE, 2002, 33 (10) :1463-1469
[7]   Measurements of the physical properties of particles in the urban atmosphere [J].
Harrison, RM ;
Jones, M ;
Collins, G .
ATMOSPHERIC ENVIRONMENT, 1999, 33 (02) :309-321
[8]  
HERING SV, 2005, UNPUB AEROSOL SCI TE
[9]  
HERING SV, 2004, IN PRESS AEROSOL SCI
[10]   Counting efficiency of condensation particle counters at low-pressures with illustrative data from the upper troposphere [J].
Hermann, M ;
Wiedensohler, A .
JOURNAL OF AEROSOL SCIENCE, 2001, 32 (08) :975-991