Heat transfer analysis and modeling of a cryogenic laser radiometer

被引:7
作者
Johnson, BC [1 ]
Kumar, AR
Zhang, ZM
Livigni, DJ
Cromer, CL
Scott, TR
机构
[1] Univ Florida, Dept Mech Engn, Gainesville, FL 32611 USA
[2] Natl Inst Stand & Technol, Div Optoelect, Sources & Detectors Grp, Boulder, CO 80303 USA
关键词
D O I
10.2514/2.6378
中图分类号
O414.1 [热力学];
学科分类号
摘要
This study investigates the laser optimized cryogenic radiometer (LOCR) recently acquired by the National Institute of Standards and Technology in Boulder, Colorado, to calibrate laser power meters and detectors. The objectives are to evaluate potentially significant sources of uncertainty in the radiometric measurements and to develop transient models that efficiently and accurately predict the behavior of this radiometer. The analysis suggests that radiation from the Brewster window assembly may cause the total power entering the radiometer to drift more than 130 nW for a room temperature variation of 0.2 K, Steady-state modeling of the LOCR with finite element analysis software indicates a relative inequivalence between optical and electrical heating of 4 x 10(-6) at the 1-mW power level. A new thermal model has been developed to simplify transient predictions by combining lumped parameter and one-dimensional elements. This model outperforms single-time-constant exponential models and can be expanded to simulate the complete radiometer system.
引用
收藏
页码:575 / 581
页数:7
相关论文
共 22 条
[1]  
*ANSYS INC, 1996, ANSYS VERS 5 3
[2]   CHARACTERIZATION OF AN ABSOLUTE CRYOGENIC RADIOMETER AS A STANDARD DETECTOR FOR RADIANT-POWER MEASUREMENTS [J].
DATLA, RU ;
STOCK, K ;
PARR, AC ;
HOYT, CC ;
MILLER, PJ ;
FOUKAL, PV .
APPLIED OPTICS, 1992, 31 (34) :7219-7225
[3]   THERMAL-CONDUCTIVITY OF NB-TI ALLOY IN LOW-TEMPERATURE RANGE [J].
FLACHBART, K ;
FEHER, A ;
JANOS, S ;
MALEK, Z ;
RYSKA, A .
PHYSICA STATUS SOLIDI B-BASIC SOLID STATE PHYSICS, 1978, 85 (02) :545-551
[4]   HISTORY OF SOLAR RADIOMETRY AND THE WORLD RADIOMETRIC REFERENCE [J].
FROHLICH, C .
METROLOGIA, 1991, 28 (03) :111-115
[5]   National Institute of Standards and Technology high-accuracy cryogenic radiometer [J].
Gentile, TR ;
Houston, JM ;
Hardis, JE ;
Cromer, CL ;
Parr, AC .
APPLIED OPTICS, 1996, 35 (07) :1056-1068
[6]  
GINNINGS DC, 1972, TEMP ITS MEAS, V4, P339
[7]  
Jensen J., 1980, BROOKHAVEN NATL LAB, V1
[8]  
Jensen J. E., 1980, BROOKHAVEN NATL LAB, V2
[9]  
JOHNSON BC, 1997, THESIS U FLORIDA GAI
[10]  
LIVIGNI DJ, 1997, ABS RAD REM SENS C O