Constellation Technology Corporation has developed a pocket-sized radiation intensity counter (RIC) for monitoring ambient radiation levels, search activities, and customs inspection. The RIC incorporates a sensitive 2.5 cm x 2.5 cm x 0.5 cm mercuric iodide (HgI2) detector operating at roam temperature. The instrument provides the user with a visual indication of the gamma-ray count-rate by means of a 7 segment bargraph scale arranged in a semi-logarithmic scale. An alarm alerts the user when the count fate exceeds a pre-selected level. Power is supplied by two commonly available AA alkaline batteries. Total power consumption <2 mW and the battery life is estimated to be >3 months under continuous use. The design of the RIC is described in detail and results of testing sensitivity, positional response, power consumption, and temperature stability are presented. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:227 / 231
页数:5
相关论文
共 7 条
[1]  
GERRISH A, 1997, P C HIGH EN BACKGR S, P57
[2]  
Gerrish V. M., 1995, SEMICONDUCTORS SEMIM, P493
[3]  
GERRISH VM, 1993, MATER RES SOC SYMP P, V302, P129, DOI 10.1557/PROC-302-129
[4]   LOW-NOISE, LOW-POWER, MINIATURIZED PROCESSING ELECTRONICS FOR X-RAY-DETECTORS [J].
IWANCZYK, JS .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 322 (03) :543-550
[5]  
IWANCZYK JS, 1993, MATER RES SOC SYMP P, V302, P79, DOI 10.1557/PROC-302-79
[6]  
KNOLL G, 1989, RAD DETECTION MEASUR, P60
[7]   PLATELET-ACTIVATING FACTOR ACTIVATES HIV PROMOTER IN TRANSFECTED SH-SY5Y NEUROBLASTOMA-CELLS AND MOLT-4 LYMPHOCYTES-T [J].
SQUINTO, SP ;
BRAQUET, P ;
BLOCK, AL ;
BAZAN, NG .
JOURNAL OF MOLECULAR NEUROSCIENCE, 1990, 2 (02) :79-84