Human cardiac high-energy phosphate metabolite concentrations by 1D-resolved NMR spectroscopy

被引:72
作者
Bottomley, PA [1 ]
Atalar, E [1 ]
Weiss, RG [1 ]
机构
[1] JOHNS HOPKINS UNIV,SCH MED,DEPT MED,DIV CARDIOL,BALTIMORE,MD 21287
关键词
heart; metabolite; quantification; phosphorus;
D O I
10.1002/mrm.1910350507
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We have developed a method that can measure high-energy phosphate metabolite concentrations in humans with 1D resolved surface-coil NMR spectroscopy. The metabolites are measured by phosphorus (P-31) NMR spectroscopy and the tissue water proton (H-1) resonance from the same volume serves as an internal concentration reference, The method requires only the additional acquisition of a H-1 data set, and a simple calibration, performed separately, to determine the ratio of the signal per proton to the signal per phosphorus nucleus. The quantification method is particularly useful for human cardiac spectroscopy, where it eliminates image-based tissue volumetry and the corrections for signal sensitivity and phase nonuniformity necessary in prior approaches. Corrections are introduced to account for blood and fat contributions to the spectra. The method was validated on phantoms of phosphate of varying concentrations and on the human calf muscle. In calf, the adenosine triphosphate (ATP) and phosphocreatine (PCr) concentrations were 5.6 +/- 1.6 (mean +/- SD) and 26 +/- 4 mmol/kg wet wt, respectively. In normal heart, [ATP] was 5.8 +/- 1.6 and [PCr] was 10 +/- 2 mmol/kg wet wt. These values are in excellent agreement with prior NMR studies and biopsy data. The protocol is easily accommodated within existing 1D cardiac patient protocols, and the same approach is advantageous for eliminating tissue volumetry and sensitivity corrections when measuring concentrations by 2D and 3D resolved spectroscopy.
引用
收藏
页码:664 / 670
页数:7
相关论文
共 29 条
[1]   ADVANCES IN PHYSIOLOGICAL CHEMISTRY BY IN-VIVO NMR - A WORKSHOP SPONSORED BY THE SOCIETY OF MAGNETIC-RESONANCE HELD IN WOODS-HOLE, MASSACHUSETTS - MARCH 22-24, 1995 [J].
BALABAN, R ;
BOTTOMLEY, P ;
BROWN, TR ;
GADIAN, D ;
MOUNTFORD, C ;
RADDA, GK ;
ROSS, BD ;
SHULMAN, RG ;
SPRINGER, C ;
UGURBIL, K .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (03) :289-292
[3]   CORRECTING HUMAN HEART P-31 NMR-SPECTRA FOR PARTIAL SATURATION - EVIDENCE THAT SATURATION FACTORS FOR PCR ATP ARE HOMOGENEOUS IN NORMAL AND DISEASE STATES [J].
BOTTOMLEY, PA ;
HARDY, CJ ;
WEISS, RG .
JOURNAL OF MAGNETIC RESONANCE, 1991, 95 (02) :341-355
[4]   PHOSPHATE METABOLITE IMAGING AND CONCENTRATION MEASUREMENTS IN HUMAN HEART BY NUCLEAR-MAGNETIC-RESONANCE [J].
BOTTOMLEY, PA ;
HARDY, CJ ;
ROEMER, PB .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (03) :425-434
[5]   MR SPECTROSCOPY OF THE HUMAN HEART - THE STATUS AND THE CHALLENGES [J].
BOTTOMLEY, PA .
RADIOLOGY, 1994, 191 (03) :593-612
[6]   OPTIMUM FLIP-ANGLES FOR EXCITING NMR WITH UNCERTAIN T-1 VALUES [J].
BOTTOMLEY, PA ;
OUWERKERK, R .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (01) :137-141
[7]  
BOTTOMLEY PA, 1989, CLIN CHEM, V35, P392
[8]   ALTERED PHOSPHATE-METABOLISM IN MYOCARDIAL-INFARCTION - P-31 MR SPECTROSCOPY [J].
BOTTOMLEY, PA ;
HERFKENS, RJ ;
SMITH, LS ;
BASHORE, TM .
RADIOLOGY, 1987, 165 (03) :703-707
[9]   COMPARISON OF CALIBRATION STRATEGIES FOR THE IN-VIVO DETERMINATION OF ABSOLUTE METABOLITE CONCENTRATIONS IN THE HUMAN BRAIN BY P-31 MRS [J].
BUCHLI, R ;
MARTIN, E ;
BOESIGER, P .
NMR IN BIOMEDICINE, 1994, 7 (05) :225-230
[10]   ASSESSMENT OF ABSOLUTE METABOLITE CONCENTRATIONS IN HUMAN TISSUE BY P-31 MRS IN-VIVO .2. MUSCLE, LIVER, KIDNEY [J].
BUCHLI, R ;
MEIER, D ;
MARTIN, E ;
BOESIGER, P .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (04) :453-458