Size, shape, and composition of wear particles from metal-metal hip simulator testing: Effects of alloy and number of loading cycles

被引:87
作者
Catelas, I
Bobyn, JD
Medley, JB
Krygier, JJ
Zukor, DJ
Huk, OL
机构
[1] McGill Univ, Dept Biomed Engn, Montreal, PQ H3A 2B4, Canada
[2] Montreal Gen Hosp, Jo Miller Orthopaed Res Lab, Montreal, PQ H3G 1A4, Canada
[3] Univ Waterloo, Dept Mech Engn, Waterloo, ON N2L 3G1, Canada
[4] McGill Univ, Lady Davis Inst Med Res, Sir Mortimer B Davis Jewish Hosp, Div Orthopaed Surg, Montreal, PQ H3T 1E2, Canada
关键词
metal-metal wear particles; characterization; hip simulator; cobalt-chromium alloy;
D O I
10.1002/jbm.a.10088
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
There has been a revived interest in metal-metal total hip replacements because of their potential for improved wear performance compared with conventional metal-polyethylene implants. The aim of the present study was to characterize metal wear particles isolated from metal-metal hip simulator testing of various clinically relevant alloys and to analyze the effects of these alloys and the number of loading cycles on wear particle characteristics. Implants were manufactured using medical-grade cobalt-chromium-molybdenum (CoCrMo) alloys that were high-carbon wrought, low-carbon wrought, or cast (with solution annealing). Testing was performed in a MATCO orbital bearing hip simulator in 95% bovine calf serum. The wear particles were isolated from the serum at test periods of 0-0.25 million cycles (Mc) (run-in wear) and 1.75-2 Mc (steady-state wear) using an enzymatic protocol previously optimized to minimize particle changes due to reagents. Isolated particles embedded in epoxy resin were characterized by transmission electron microscopy (TEM) and energy dispersive X-ray analysis (EDXA). The EDXA results revealed the predominance of "lighter" particles containing Cr and 0 (most likely chromium oxide particles from the passivation layer) and fewer darker CoCrMo particles, with varying ratios of Co and Cr (possibly from carbides and from implant matrix material). More CoCrMo particles were observed with the low-carbon wrought alloy, but the majority of the particles for all three alloys was chromium oxides, especially for the 1.75-2 Mc test period. Image analysis of TEM micrographs revealed that for 0 - 0.25 Mc, there was up to 21% needle-shaped particles but that the majority remained round to oval in shape, reflecting the predominance of chromium oxide particles. Particle length averaged about 52 +/- 4 run, with only small differences due to the alloy. For 1.75-2 Mc, most particles were round to oval in shape. They were even less needle-shaped than at 0.25 Mc, and they had a slightly smaller length, averaging 46 +/- 3 nm. In addition to characterizing the size and shape of particles from a MATCO simulator, this study is the first to demonstrate that particles that do not contain Co (presumably chromium oxides) can be predominant in the wear of metal-metal hip implants. It is therefore recommended that future in vitro and in vivo studies include the effects of these particles rather than just the effects of CoCrMo particles on the overall tissue response. (C) 2003 Wiley Periodicals, Inc.
引用
收藏
页码:312 / 327
页数:16
相关论文
共 42 条
[1]
Amstutz H. C., 1996, CLIN ORTHOP S, V329, P11
[2]
TISSUE DIGESTION FOR WEAR DEBRIS PARTICLE ISOLATION [J].
CAMPBELL, P ;
MA, S ;
SCHMALZRIED, T ;
AMSTUTZ, HC .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1994, 28 (04) :523-526
[3]
Metal-on-metal hip replacements: Wear performance and cellular response to wear particles [J].
Campbell, P ;
McKellop, H ;
Alim, R ;
Mirra, J ;
Nutt, S ;
Dorr, L ;
Amstutz, HC .
COBALT-BASE ALLOYS FOR BIOMEDICAL APPLICATIONS, 1999, 1365 :193-209
[4]
CAMPBELL PA, 1994, T ORTHOP RES SOC, V19, P848
[5]
Catelas I, 2001, J BIOMED MATER RES, V55, P330, DOI 10.1002/1097-4636(20010605)55:3<330::AID-JBM1021>3.0.CO
[6]
2-Y
[7]
Catelas I, 2001, J BIOMED MATER RES, V55, P320, DOI 10.1002/1097-4636(20010605)55:3<320::AID-JBM1020>3.0.CO
[8]
2-3
[9]
CATELAS I, 1999, T CAN ORTHOP RES SOC, P43
[10]
CATELAS I, 2000, T 6 WORLD BIOM C 200, P482