Selective determination of gas dialysable components in complex sample solutions using triangle programmed coulometric titration in continuous flow systems.

被引:12
作者
Becker, M [1 ]
Fuhrmann, B [1 ]
Spohn, U [1 ]
机构
[1] UNIV HALLE WITTENBERG,INST BIOTECHNOL,D-06120 HALLE,GERMANY
关键词
ammonium/ammonia; hydrogen sulphide/sulphide; gas dialysis; coulometry; chemiluminescence; flow system;
D O I
10.1016/0003-2670(96)00015-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A precisely and sensitively working coulometric flow titration procedure was developed to determine ammonium/ammonia and sulphide/hydrogen sulphide in leachate from a municipal waste landfill site. The selectivity of coulometric titration with electrogenerated hypobromite could be essentially improved by the combination with a continuously and quantitatively working thin layer gas dialysis cell, which was integrated in the flow titration setup in one measuring device. During the flow titrations both ammonium/ammonia and sulphide/hydrogen sulphide were quantitatively separated from the sample solutions. Both the commonly used biamperometric and the chemiluminescence flow detection could be applied to follow the titration course. By triangle-programmed coulometric titration ammonium/ammonia were detected in a linear range from 10(-3) to 10(-6) M (relative standard deviation (RSD) = 1%, n = 6, confidence level alpha = 0.05) using biamperometric detection and from 5 x 10(-4) to 10(-6) M (RSD = 3%, n = 6, alpha = 0.05) using chemiluminescence detection. Sulphide/hydrogen sulphide could be determined in a linear range from 5 . 10(-4) to 2 . 10(-6) M (RSD = 4%, n = 6,alpha = 0.05). Up to 30 flow titrations could be implemented per hour.
引用
收藏
页码:115 / 123
页数:9
相关论文
共 33 条
[1]  
AGASYAN LB, 1966, VEST MOSK U 2 K, V21, P96
[2]   ULTRAMICROCOULOMETRIC DETERMINATION OF AMMONIA IN SERUM, PLASMA, AND BLOOD [J].
ATTIYAT, AS .
MICROCHEMICAL JOURNAL, 1987, 35 (03) :340-346
[4]   COULOMETRIC TITRATION OF AMMONIA WITH HYPOBROMITE USING DIRECT AMPEROMETRIC END POINT DETECTION [J].
CHRISTIAN, GD ;
PURDY, WC ;
KNOBLOCK, EC .
ANALYTICAL CHEMISTRY, 1963, 35 (13) :2217-&
[5]  
COOK C, 1989, TEX J SCI, V41, P273
[6]  
Fang Z., 1993, FLOW INJECTION SEPAR
[7]   MEMBRANE BASED GAS SAMPLING AND ANALYSIS COUPLED TO CONTINUOUS-FLOW SYSTEMS [J].
FRENZEL, W .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 342 (10) :817-821
[8]  
FRITSCHE U, 1980, MIKROCHIM ACTA, V2, P85
[9]   A PC-BASED TITRATOR FOR FLOW GRADIENT TITRATIONS [J].
FUHRMANN, B ;
SPOHN, U .
JOURNAL OF AUTOMATIC CHEMISTRY, 1993, 15 (06) :209-216
[10]   VOLUMETRIC TRIANGLE-PROGRAMMED FLOW TITRATIONS BASED ON PRECISELY GENERATED CONCENTRATION GRADIENTS [J].
FUHRMANN, B ;
SPOHN, U .
ANALYTICA CHIMICA ACTA, 1993, 282 (02) :397-406