Development of a high-throughput automated analyzer using biochip array technology

被引:137
作者
FitzGerald, SP [1 ]
Lamont, JV [1 ]
McConnell, RI [1 ]
Benchikh, EO [1 ]
机构
[1] Randox Labs, Crumlin BT29 4QY, Antrim, North Ireland
关键词
D O I
10.1373/clinchem.2005.049429
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Use of protein array technology over conventional assay methods has advantages that include simultaneous detection of multiple analytes, reduction in sample and reagent volumes, and high output of test results. The susceptibility of ligands to denaturation, however, has impeded production of a stable, reproducible biochip platform, limiting most array assays to manual or, at most, semiautomated processing techniques. Such limitations may be overcome by novel biochip fabrication procedures. Methods: After selection of a suitable biochip substrate, biochip surfaces were chemically modified and assessed to enable optimization of biochip fabrication procedures for different test panels. The assay procedure was then automated on a dedicated instrument, and assay performance was determined for a panel of cytokine markers. Assay results were then compared with a commercial method for measurement of cytokine markers. Results: Secondary ion mass spectrometry and x-ray photoelectron spectroscopy demonstrated appropriate and reproducible modification of the biochip surface. Contact-angle studies also confirmed generation of hydrophobic surfaces that enabled containment of droplets for fabrication of discrete test regions. Automation of the biochip assays on a dedicated instrument produced excellent cytokine marker performance with intra- and interassay imprecision < 10% for most analytes. Comparison studies showed good agreement with other methods (r = 0.95-0.99) for cytokines. Conclusion: Performance data from this automated biochip array analyzer provide evidence that it is now possible to produce stable and reproducible biochips for output of more than 2000 test results per hour. (c) 2005 American Association for Clinical Chemistry.
引用
收藏
页码:1165 / 1176
页数:12
相关论文
共 17 条
[1]   Comparison of antibody functionality using different immobilization methods [J].
Danczyk, R ;
Krieder, B ;
North, A ;
Webster, T ;
HogenEsch, H ;
Rundell, A .
BIOTECHNOLOGY AND BIOENGINEERING, 2003, 84 (02) :215-223
[2]   MEASUREMENT OF FREE HORMONES IN BLOOD [J].
EKINS, R .
ENDOCRINE REVIEWS, 1990, 11 (01) :5-46
[3]  
EKINS R, 1990, ANN BIOL CLIN-PARIS, V48, P655
[4]  
Ekins RP, 1998, CLIN CHEM, V44, P2015
[5]  
FITZGERALD SP, 2002, Patent No. 6498010
[6]  
JONSSON U, 1985, BIOCHEM J, V227, P363
[7]   Protein microarrays: A literature survey [J].
Kricka, LJ ;
Joos, T ;
Fortina, P .
CLINICAL CHEMISTRY, 2003, 49 (12) :2109-2109
[8]   Solid supports for microarray immunoassays [J].
Kusnezow, W ;
Hoheisel, JD .
JOURNAL OF MOLECULAR RECOGNITION, 2003, 16 (04) :165-176
[9]   Antibody microarrays: An evaluation of production parameters [J].
Kusnezow, W ;
Jacob, A ;
Walijew, A ;
Diehl, F ;
Hoheisel, JD .
PROTEOMICS, 2003, 3 (03) :254-264
[10]   THE INFLUENCE OF ADSORPTION OF NATIVE AND MODIFIED ANTIBODIES ON THEIR ACTIVITY [J].
LIN, JN ;
ANDRADE, JD ;
CHANG, IN .
JOURNAL OF IMMUNOLOGICAL METHODS, 1989, 125 (1-2) :67-77