Elimination of the spectral interference from polyatomic ions with rare earth elements in inductively coupled plasma mass spectrometry by combining algebraic correction with chromatographic separation

被引:44
作者
Cao, XD
Yin, M
Wang, XR
机构
[1] Nanjing Univ, State Key Lab Pollut Control & Resources Reuse, Sch Environm, Nanjing 210093, Peoples R China
[2] Chinese Acad Geol Sci, Inst Rock & Mineral Anal, Beijing 100037, Peoples R China
基金
中国国家自然科学基金;
关键词
inductively coupled plasma mass spectrometry; rare earth elements; interference elimination; algebraic correction; chromatographic separation;
D O I
10.1016/S0584-8547(01)00170-7
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A simple and effective procedure is developed to avoid the spectral interference from light rare earth elements (REEs) and barium polyatomic ions on some rare earth elements in inductively coupled plasma mass spectrometry (ICP-MS) by combining algebraic correction with AG50W-x8 cation exchangeable chromatography. Algebraic correction is made to reduce the spectroscopic overlap interference of (PrO)-Pr-141-O-16 and (NdO)-Nd-143-O-16 on Gd-157 and Tb-159. The spectroscopic overlap interference of BaO+ and BaOH+ on some middle REEs are overcome by separation of REEs from barium with AG50W-x 8 cation exchangeable chromatography. Prior to the determination, REEs are separated from complicated matrix samples using AG50W-x 8 cation exchangeable resin. Ba is eluted with 2 mol/l HNO3 solution. REEs are retained and could then be eluted with 5 mol/l HNO3 solution. Recoveries for REEs are from 96 to 110%. More than 99.5% of Ba in the sample is removed,ensuring that the spectral interference from barium polyatomic ions on some middle REEs such as Nd, Sm, Eu and Gd are eliminated. The potential of the proposed method is evaluated by analysis of Certified Reference Materials (CRMs). Results show that experimental data are in good agreement with the certified values. The new technique has been successfully employed for the determination of REEs in practical soil and plant Samples. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:431 / 441
页数:11
相关论文
共 18 条
[1]   Determination of the rare earth elements in aqueous samples at sub-ppt levels by inductively coupled plasma mass spectrometry and flow injection ICPMS [J].
Aggarwal, JK ;
Shabani, MB ;
Palmer, MR ;
Ragnarsdottir, KV .
ANALYTICAL CHEMISTRY, 1996, 68 (24) :4418-4423
[2]   CORRECTION OF INTERFERENCE EFFECTS IN THE DETERMINATION OF THE RARE-EARTH ELEMENTS AND HAFNIUM BY SPARK-SOURCE MASS-SPECTROMETRY [J].
BACON, JR ;
URE, AM .
ANALYTICA CHIMICA ACTA, 1979, 105 (01) :163-176
[3]  
CAO XD, 1998, CHIN J ENV CHEM, V2, P132
[4]  
CAO XD, 2000, INT J ENVIRON AN CH, V4, P295
[5]   THE GROUP SEPARATION OF THE RARE-EARTH ELEMENTS AND YTTRIUM FROM GEOLOGIC MATERIALS BY CATION-EXCHANGE CHROMATOGRAPHY [J].
CROCK, JG ;
LICHTE, FE ;
WILDEMAN, TR .
CHEMICAL GEOLOGY, 1984, 45 (1-2) :149-163
[6]   SPECTRAL AND NON-SPECTRAL INTERFERENCES IN INDUCTIVELY-COUPLED PLASMA MASS-SPECTROMETRY [J].
DAMS, RFJ ;
GOOSSENS, J ;
MOENS, L .
MIKROCHIMICA ACTA, 1995, 119 (3-4) :277-286
[7]   DETERMINATION OF RARE-EARTH ELEMENTS IN GEOLOGICAL SAMPLES BY INDUCTIVELY COUPLED PLASMA SOURCE-MASS SPECTROMETRY [J].
DATE, AR ;
HUTCHISON, D .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (03) :269-276
[8]   RARE-EARTH ELEMENT GEOCHEMISTRY OF STANDARD SEDIMENTS - A STUDY USING INDUCTIVELY COUPLED PLASMA SPECTROMETRY [J].
JARVIS, I ;
JARVIS, KE .
CHEMICAL GEOLOGY, 1985, 53 (3-4) :335-344
[9]   AVOIDANCE OF SPECTRAL INTERFERENCE ON EUROPIUM IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY BY SENSITIVE MEASUREMENT OF THE DOUBLY CHARGED ION [J].
JARVIS, KE ;
GRAY, AL ;
MCCURDY, E .