MRI of the neonatal brain: Optimization of spin-echo parameters

被引:36
作者
Jones, RA [1 ]
Palasis, S [1 ]
Grattan-Smith, JD [1 ]
机构
[1] Emory Univ, Sch Med, Dept Radiol, Atlanta, GA 30322 USA
关键词
D O I
10.2214/ajr.182.2.1820367
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. Our aim was to measure the relaxation times of the neonatal brain and to use these to derive pulse sequence parameters that enhance the signal-to-noise ratio (SNR) and contrast of MRI scans of the neonatal brain. SUBJECTS AND METHODS. The transverse (T2) and longitudinal (T1) relaxation times were measured for 10 healthy neonates, and the average relaxation times were calculated for both gray and white matter. Simulations using these values were then performed to estimate the optimal pulse sequence parameters. Images were obtained in three neonates using both the optimized and conventional sequence parameters. RESULTS. The measured (mean +/- SD) relaxation times of the neonatal brain at 1.5 T were T1 equals 1,712 +/- 235 msec and T2 equals 394 +/- 52 msec in white matter and T1 equals 1,144 +/- 245 msec and T2 equals 206 +/- 26 msec in gray matter. The optimized T1-weighted imaging used a turbo spin-echo sequence with an echo-train length of 3 and TR/TE of 850/11 msec and showed increases in both the contrast and the SNR. The optimized T2-weighted sequence used a TE of 270 msec and markedly increased the contrast but at the expense of a reduction in the SNR. CONCLUSION. Parameters of MRI turbo spin-echo sequences for scanning neonates are different from those required for adult studies, and appropriate protocols should be used.
引用
收藏
页码:367 / 372
页数:6
相关论文
共 8 条
  • [1] BARKOVICH AJ, 2000, PEDIAT NEUROIMAGING, P4
  • [2] QUANTITATIVE GROWTH AND DEVELOPMENT OF HUMAN BRAIN
    DOBBING, J
    SANDS, J
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD, 1973, 48 (10) : 757 - 767
  • [3] NUCLEAR-RELAXATION OF HUMAN BRAIN GRAY AND WHITE MATTER - ANALYSIS OF FIELD-DEPENDENCE AND IMPLICATIONS FOR MRI
    FISCHER, HW
    RINCK, PA
    VANHAVERBEKE, Y
    MULLER, RN
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 16 (02) : 317 - 334
  • [4] THE RICIAN DISTRIBUTION OF NOISY MRI DATA
    GUDBJARTSSON, H
    PATZ, S
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (06) : 910 - 914
  • [5] RELAXOMETRY OF BRAIN - WHY WHITE MATTER APPEARS BRIGHT IN MRI
    KOENIG, SH
    BROWN, RD
    SPILLER, M
    LUNDBOM, N
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (03) : 482 - 495
  • [6] IMMATURE BRAIN - SPIN-ECHO PULSE SEQUENCE PARAMETERS FOR HIGH-CONTRAST MR IMAGING
    NOWELL, MA
    HACKNEY, DB
    ZIMMERMAN, RA
    BILANIUK, LT
    GROSSMAN, RI
    GOLDBERG, HI
    [J]. RADIOLOGY, 1987, 162 (01) : 272 - 273
  • [7] CT versus MR in neonatal brain imaging at term
    Robertson, RL
    Robson, CD
    Zurakowski, D
    Antiles, S
    Strauss, K
    Mulkern, RV
    [J]. PEDIATRIC RADIOLOGY, 2003, 33 (07) : 442 - 449
  • [8] WEHRLI FW, 1991, FAST SCAN MAGNETIC R, P34