Count rate meter with a time constant using adaptive digital signal processing

被引:1
作者
Zigic, A [1 ]
Arandjelovic, V [1 ]
机构
[1] VINCA, Dept Elect, Inst Nucl Sci, Belgrade 11001, Serbia Monteneg
关键词
time constant count rate meter; adaptive signal processing; low-pass filter; FIR filter; mean count rate;
D O I
10.1016/j.radmeas.2005.08.001
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Two methods developed to improve the classical time constant count rate meters by using the adaptive signal-processing tools are presented. An optimized detection algorithm that senses the change of the mean count rate was implemented in both methods. Three different types of low-pass filters of various structures with adaptive parameters to implement the control of the mean count rate error by suppressing the fluctuations in a controllable way were considered, and one of them was implemented in both methods. An adaptation algorithm for time constant interval calculation was executed after the low-pass filter was devised and implemented in the first method. This adaptation algorithm makes it possible to obtain shorter time constant intervals for higher stationary mean count rates. The adaptation algorithm for time constant interval calculation executed before the low-pass filter was devised and implemented in the second method. This adaptation algorithm enables sensing of the rapid change of the mean count rate before the fluctuations suppression is carried out. Some parameters were fixed to their optimum values after appropriate optimization procedures had been performed. The implemented low-pass filter has variable number of stationary coefficients depending on the specified error and the mean count rate. It implements the control of the mean count rate error by suppressing the fluctuations in a controllable way. The simulated and realized methods, using the developed algorithms, guarantee: a response time not in excess of 2 s for a mean count rate higher than 2 counts/s and a controllable mean count rate error in the range of 4-10%. (c) 2005 Elsevier Ltd. All rights reserved.
引用
收藏
页码:295 / 305
页数:11
相关论文
共 13 条
[1]  
Antoniou A., 1979, DIGITAL FILTERS ANAL
[2]   A software method for suppressing statistical fluctuations in preset count digital-rate meter algorithms [J].
Arandjelovic, V ;
Koturovic, A ;
Vukanovic, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (05) :2561-2566
[3]   The optimum dynamic parameters of digital rate meter algorithms [J].
Arandjelovic, VD ;
Koturovic, AM .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (12) :4609-4611
[4]   The optimum dynamics of preset count digital rate meter algorithms [J].
Arandjelovic, VD ;
Koturovic, AM .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (12) :4142-4145
[5]   The dynamic properties of the preset count digital rate meter algorithms [J].
Arandjelovic, VD ;
Koturovic, AM ;
Vukanovic, RB .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (08) :2902-2908
[6]   COMPARING A RECURSIVE DIGITAL-FILTER WITH THE MOVING-AVERAGE AND SEQUENTIAL PROBABILITY-RATIO DETECTION METHODS FOR SNM PORTAL MONITORS [J].
FEHLAU, PE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (02) :143-146
[7]  
HAYKIN S., 1986, ADAPTIVE FILTER THEO
[8]  
KNOLL GF, 1999, RAD DETECTION MEASUR, P142
[9]   ANALYSIS OF RESPONSE-TIMES FOR DIGITAL RATE METER ALGORITHMS [J].
KOTUROVIC, AM ;
ARANDJELOVIC, VD .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (05) :3374-3376
[10]  
MILIC L, 1998, ELECTRON ENG, V47, pE7