Fabrication of electrophoresis devices on quartz and glass substrates using a bonding with HF solution

被引:25
作者
Nakanishi, H
Nishimoto, T
Nakamura, N
Nagamachi, S
Arai, A
Iwata, Y
Mito, Y
机构
来源
MEMS 97, PROCEEDINGS - IEEE THE TENTH ANNUAL INTERNATIONAL WORKSHOP ON MICRO ELECTRO MECHANICAL SYSTEMS: AN INVESTIGATION OF MICRO STRUCTURES, SENSORS, ACTUATORS, MACHINES AND ROBOTS | 1997年
关键词
D O I
10.1109/MEMSYS.1997.581834
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Capillary electrophoresis (CE) devices have been fabricated in a small size on quartz or glass substrates. To fabricate a quartz device, we developed a novel bonding method with HF solution. HF-bonding was carried out with 1 % HF solution with a pressure of similar to 40 gf/cm(2). With an evaluation of the new bonding method, we confirm that the method ensures enough strength, air tightness and optical property. An advantage of this method is simplicity of bonding process in comparison with a conventional thermal bonding. Using the glass mu CE devices, we obtained the original device design and process procedure through our experiments. It was proved that mu CE chip permits to apply higher electric power in comparison with the conventional CE system. We show that our design permits to perform quantitative sample injection (several 10 pi or less), effective separation (the number of theoretical plates is over 2500) and high speed analysis (over 10 times as high as the conventional CE system). These features promise an advantage of mu CE devices over conventional analytical equipments.
引用
收藏
页码:299 / 304
页数:6
相关论文
empty
未找到相关数据