On-line column-switching high-performance liquid chromatography analysis of cardiovascular drugs in serum with automated sample clean-up and zone-cutting technique to perform chiral separation

被引:40
作者
Mangani, F [1 ]
Luck, G [1 ]
Fraudeau, C [1 ]
Verette, E [1 ]
机构
[1] GILSON MED ELECT FRANCE SA,F-95400 VILLIERS LE BEL,FRANCE
关键词
column switching; enantiomer separation; lidocaine; beta-blockers; diltiazem; verapamil;
D O I
10.1016/S0021-9673(96)00878-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A selective and highly reproducible, multi-column HPLC method is described for the analysis of the following cardiovascular drugs: lidocaine, pindolol, metoprolol, oxprenolol, diltiazem and verapamil, in serum. Column-switching devices are employed in combination with advanced separation media technologies for the automated analysis of samples containing complex matrices. The method consists of on-line sample clean-up using a restricted access sorbent, HPLC analysis of the drugs on a microsphere non-porous silica RP-18 column, and front-cutting to perform the chiral separation of pindolol enantiomers on a second HPLC system. Simultaneous control of the two HPLC systems and data analysis is achieved from a single centralized software. The R.S.D. values of the peak areas for spiked serum are less than 1% for metoprolol and oxprenolol, 2-5% for lidocaine, diltiazem and verapamil, and 1.2 and 2.4% for the two pindolol enantiomers. Recoveries, limits of detection and linearities are provided.
引用
收藏
页码:235 / 241
页数:7
相关论文
共 11 条
[1]   ALKYL-DIOL SILICA (ADS) - RESTRICTED ACCESS PRECOLUMN PACKINGS FOR DIRECT-INJECTION AND COUPLED-COLUMN CHROMATOGRAPHY OF BIOFLUIDS [J].
BOOS, KS ;
RUDOLPHI, A ;
VIELHAUER, S ;
WALFORT, A ;
LUBDA, D ;
EISENBEISS, F .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1995, 352 (7-8) :684-690
[2]   COLUMN-SWITCHING TECHNIQUES FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF DRUGS IN BIOLOGICAL SAMPLES [J].
CAMPINSFALCO, P ;
HERRAEZHERNANDEZ, R ;
SEVILLANOCABEZA, A .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1993, 619 (02) :177-190
[3]   THE CHIRAL CHROMATOGRAPHIC-SEPARATION OF BETA-ADRENOCEPTOR BLOCKING-DRUGS [J].
DYAS, AM .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1992, 10 (06) :383-404
[4]   SEPARATION OF THE ENANTIOMERS OF BETA-RECEPTOR BLOCKING-AGENTS AND OTHER CATIONIC DRUGS USING A CHIRAL-AGP COLUMN - BINDING-PROPERTIES AND CHARACTERIZATION OF IMMOBILIZED ALPHA-1-ACID GLYCOPROTEIN [J].
ENQUIST, M ;
HERMANSSON, J .
JOURNAL OF CHROMATOGRAPHY, 1990, 519 (02) :285-298
[5]   DETERMINATION OF DRUGS BY DIRECT-INJECTION OF PLASMA INTO A BIOCOMPATIBLE EXTRACTION COLUMN BASED ON A PROTEIN-ENTRAPPED HYDROPHOBIC PHASE [J].
HERMANSSON, J ;
GRAHN, A .
JOURNAL OF CHROMATOGRAPHY A, 1994, 660 (1-2) :119-129
[6]  
HERMANSSON J, 1988, CHROMATOGRAPHIC CHIR, P245
[7]  
HULPKE H, 1986, PRACTICE HIGH PERFOR, P143
[8]   Selective extraction of beta-blockers from biological fluids by column-switching high-performance liquid chromatography using an internal-surface phenylboronic acid precolumn [J].
Ohta, T ;
Niida, S ;
Nakamura, H .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1996, 675 (01) :168-173
[9]   AN IMPROVED INTERNAL SURFACE REVERSED PHASE [J].
PERRY, JA ;
INVERGO, B ;
WAGNER, H ;
SZCZERBA, TJ ;
RATEIKE, JD .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1992, 15 (18) :3343-3352
[10]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY PACKING MATERIALS FOR THE ANALYSIS OF SMALL MOLECULES IN BIOLOGICAL MATRICES BY DIRECT INJECTION [J].
PINKERTON, TC .
JOURNAL OF CHROMATOGRAPHY, 1991, 544 (1-2) :13-23