Determining in vivo biomechanical properties of the cornea with an ocular response analyzer

被引:1008
作者
Luce, DA [1 ]
机构
[1] Reichert Inc, Depew, NY USA
关键词
D O I
10.1016/j.jcrs.2004.10.044
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To study the results of an ocular response analyzer (ORA) to determine the biomechanical properties of the cornea and their relationship to intraocular pressure (lop). Setting: Reichert Inc., Depew, New York, USA. Methods: The ORA (Reichert) makes 2 essentially instantaneous applanation measurements that permit determination of corneal and IOP effects. Results: Measurements of several populations indicate that corneal hysteresis, a biomechanical measure, varied over a dynamic range of 1.8 to 14.6 mm Hg and was only weakly correlated with corneal thickness (r(2) = 0.12); this is related to the observation that some subjects with relatively thick corneas have less-than-average corneal hysteresis. Corneal hysteresis changes diurnally, presumably as a result of hydration changes. Keratoconus, Fuchs' dystrophy, and post-LASIK patients demonstrated low corneal hysteresis. Conclusion: The corneal hysteresis biornechanical measure may prove valuable for qualification and predictions of outcomes of refractive surgery and in other cases in which corneal biomechanics are important.
引用
收藏
页码:156 / 162
页数:7
相关论文
共 3 条
[1]   Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach [J].
Doughty, MJ ;
Zaman, ML .
SURVEY OF OPHTHALMOLOGY, 2000, 44 (05) :367-408
[2]  
Luce DA, 2002, US patent, Patent No. [419 631, 419631]
[3]  
Soergel F, 1995, Ger J Ophthalmol, V4, P151