Radiation exposure to children during cardiac catheterization

被引:61
作者
Boothroyd, A
McDonald, E
Moores, BM
Sluming, V
Carty, H
机构
[1] ROYAL LIVERPOOL CHILDRENS NHS TRUST,DEPT RADIOL,LIVERPOOL L12 2AP,MERSEYSIDE,ENGLAND
[2] INTEGRATED RADIOL SERV,UNIT 188,LIVERPOOL L3 4BJ,MERSEYSIDE,ENGLAND
[3] UNIV LIVERPOOL,DEPT RADIOG,LIVERPOOL L69 3BX,MERSEYSIDE,ENGLAND
关键词
D O I
10.1259/bjr.70.830.9135445
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to assess the suitability of different radiation dosimetry methods and record radiation exposures during paediatric catheterization. Three methods of dosimetry were employed: thermoluminescent dosimetry (TLD), dose-area product and calculation of entrance surface dose from calibrated exposure factors. Examinations included bi-plane fluoroscopy, and cineangiography for diagnosis and treatment of congenital heart disease. The most suitable method of radiation dosimetry for cardiac catheterization is the use of calculated entrance dose or a dose-area product meter. Children were exposed to high levels of radiation during cardiac catheterization but there was a wide variation in radiation dosage. Careful consideration should be given to the suitability of radiation dosimetry for cardiac catheterization.
引用
收藏
页码:180 / 185
页数:6
相关论文
共 17 条
[1]   GUIDELINES FOR PEDIATRIC THERAPEUTIC CARDIAC-CATHETERIZATION - A STATEMENT FOR HEALTH-PROFESSIONALS FROM THE COMMITTEE ON CONGENITAL CARDIAC DEFECTS OF THE COUNCIL ON CARDIOVASCULAR-DISEASE IN THE YOUNG, THE AMERICAN-HEART-ASSOCIATION [J].
ALLEN, HD ;
DRISCOLL, DJ ;
FRICKER, FJ ;
HERNDON, P ;
MULLINS, CE ;
SNIDER, AR ;
TAUBERT, KA .
CIRCULATION, 1991, 84 (05) :2248-2258
[2]  
AMIEL M, 1975, ANN RADIOL, V19, P623
[3]   EPIDEMIOLOGIC EVALUATION OF RADIATION RISK USING POPULATIONS EXPOSED AT HIGH-DOSES [J].
DARBY, SC .
HEALTH PHYSICS, 1986, 51 (03) :269-281
[4]   RADIATION-DOSES AND SOMATIC RISK TO PATIENTS DURING CARDIAC RADIOLOGICAL PROCEDURES [J].
FAULKNER, K ;
LOVE, HG ;
SWEENEY, JK ;
BARDSLEY, RA .
BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (700) :359-363
[5]   ANGLED VIEWS IN CINEANGIOCARDIOGRAPHY OF CONGENITAL HEART-DISEASE [J].
FELLOWS, KE ;
KEANE, JF ;
FREED, MD .
CIRCULATION, 1977, 56 (03) :485-490
[6]  
GUSTAFSSON M, 1976, BRIT J RADIOL, V49, P686, DOI 10.1259/0007-1285-49-584-686
[7]   ABSORBED, GONAD AND INTEGRAL DOSES TO PATIENT AND PERSONNEL FROM ANGIOGRAPHIC PROCEDURES [J].
KAUDE, J ;
SVAHN, G .
ACTA RADIOLOGICA-DIAGNOSIS, 1974, 15 (04) :454-464
[8]   ASSESSMENT OF LITHIUM BORATE THERMOLUMINESCENT DOSIMETERS FOR MEASUREMENT OF DOSES TO PATIENTS IN DIAGNOSTIC RADIOLOGY [J].
LANGMEAD, WA ;
WALL, BF ;
PALMER, KE .
BRITISH JOURNAL OF RADIOLOGY, 1976, 49 (587) :956-962
[9]  
MARIPUU E, 1974, 017 NAT I RAD PROT
[10]   RADIATION EXPOSURE TO THE PEDIATRIC-PATIENT FROM CARDIAC-CATHETERIZATION AND ANGIOCARDIOGRAPHY [J].
MARTIN, EC ;
OLSON, A .
BRITISH JOURNAL OF RADIOLOGY, 1980, 53 (626) :100-106