Initial stability of a cementless acetabular cup design: Experimental investigation on the effect of adding fins to the rim of the cup

被引:52
作者
Baleani, M
Fognani, R
Toni, A
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Ist Ortoped Rizzoli, Div Ortopedia & Traumatol 1, I-40136 Bologna, Italy
关键词
hip arthroplasty; acetabular cup; cementless fixation; initial stability;
D O I
10.1046/j.1525-1594.2001.025008664.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Different design solutions have been suggested for improvement of the initial stability of cementless acetabular cups, such as adding threads, spikes, or pegs to the hemispherical geometry, the pore structure of the surface. and screw fixation. This experimental study investigated the effect of fins on the initial stability of the acetabular cup. Three designs were studied, with none, 2, and 12 fins, respectively. The cups were press fit into cavities reamed in 2 different polyurethane foams, used to simulate 2 qualities of cancellous bone. Two millimeter press-fit and exact-fit conditions were investigated. The results show that the type of substrate and the interference value are important in determining the initial stability of the cup. The addition of fins on the cup rim enhances in vitro the initial stability, especially in cases of a poor press fit with a good substrate. This preclinical investigation suggests that the use of a cup design with fins may be beneficial in all cases in which press fit of the cup cannot be assured. However, further clinical studies are required to validate in vivo the efficacy of the fins as additional fixation devices.
引用
收藏
页码:664 / 669
页数:6
相关论文
共 26 条
[1]
Adler E, 1992, J Arthroplasty, V7, P295, DOI 10.1016/0883-5403(92)90052-R
[2]
[Anonymous], 67 ANN M AM AC ORTH
[3]
IMPLANT FIXATION IMPROVED BY CLOSE FIT - CYLINDRICAL IMPLANT - BONE INTERFACE STUDIED IN RABBITS [J].
CARLSSON, L ;
ROSTLUND, T ;
ALBREKTSSON, B ;
ALBREKTSSON, T .
ACTA ORTHOPAEDICA SCANDINAVICA, 1988, 59 (03) :272-275
[4]
Clarke H J, 1991, J Arthroplasty, V6, P335, DOI 10.1016/S0883-5403(06)80185-3
[5]
COLLIER JP, 1988, CLIN ORTHOP RELAT R, V235, P173
[6]
THE INITIAL STABILITY OF UNCEMENTED ACETABULAR COMPONENTS [J].
CURTIS, MJ ;
JINNAH, RH ;
WILSON, VD ;
HUNGERFORD, DS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (03) :372-376
[7]
Dammak M, 1997, J BIOMED MATER RES, V35, P329, DOI 10.1002/(SICI)1097-4636(19970605)35:3<329::AID-JBM7>3.0.CO
[8]
2-J
[9]
HAVELIN LI, 2000, 67 ANN M AM AC ORTH
[10]
JASTY M, 1994, CLIN ORTHOP RELAT R, P111