3D spatial compounding of ultrasound images using image-based nonrigid registration

被引:60
作者
Krücker, JF [1 ]
Meyer, CR [1 ]
LeCarpentier, GL [1 ]
Fowlkes, JB [1 ]
Carson, PL [1 ]
机构
[1] Univ Michigan, Dept Radiol, Ann Arbor, MI 48109 USA
关键词
ultrasound imaging; ultrasonics; medical imaging; 3D ultrasound; 3D spatial compounding; automatic registration; mutual information; thin plate spline; refraction correction;
D O I
10.1016/S0301-5629(00)00286-6
中图分类号
O42 [声学];
学科分类号
070206 [声学]; 082403 [水声工程];
摘要
Medical ultrasound images are often distorted enough to significantly limit resolution during compounding (i.e., summation of images from multiple views). A new, volumetric image registration technique has been used successfully to enable high spatial resolution in three-dimensional (3D) spatial compounding of ultrasound images. Volumetric ultrasound data were acquired by scanning a linear matrix array probe in the elevational direction in a focal lesion phantom and in a breast in vitro. To obtain partly uncorrelated views, the volume of interest was scanned at five different transducer tilt angles separated by 4 degrees to 6 degrees. Pairs of separate views were registered by an automatic procedure based on a mutual information metric, using global full affine and thin-plate spline warping transformations. Registration accuracy was analyzed automatically in the phantom data, and manually in vivo, yielding average registration errors of 0.31 mm and 0.65 mm, respectively. In the vicinity of the warping control points, registrations obtained with warping transformations were significantly more accurate than full affine registrations. Compounded images displayed the expected reduction in speckle noise and increase in contrast-to-noise ratio (CNR), as well as better delineation of connective tissues and reduced shadowing. Compounding also revealed some apparent low contrast lobulations that were not visible in the single-sweep images. Given expected algorithmic and hardware enhancements, nonrigid, image-based registration shows great promise for reducing tissue motion and refraction artifacts in 3D spatial compounding. (C) 2001 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:1475 / 1488
页数:14
相关论文
共 31 条
[1]
Three-dimensional freehand ultrasound: Image reconstruction and volume analysis [J].
Barry, CD ;
Allott, CP ;
John, NW ;
Mellor, PM ;
Arundel, PA ;
Thomson, DS ;
Waterton, JC .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (08) :1209-1224
[2]
COMPOUND SCANNING WITH AN ELECTRICALLY STEERED BEAM [J].
BERSON, M ;
RONCIN, A ;
POURCELOT, L .
ULTRASONIC IMAGING, 1981, 3 (03) :303-308
[3]
SPECKLE IN ULTRASOUND B-MODE SCANS [J].
BURCKHARDT, CB .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1978, 25 (01) :1-6
[4]
BUTTERY B, 1984, J ULTRAS MED, V3, P49
[5]
Fenn RC, 1997, ACOUST IMAG, V23, P237
[6]
HOLMES J H, 1954, Trans Am Clin Climatol Assoc, V66, P208
[7]
Multi-angle compound imaging [J].
Jespersen, SK ;
Wilhjelm, JE ;
Sillesen, H .
ULTRASONIC IMAGING, 1998, 20 (02) :81-102
[8]
SNAKES - ACTIVE CONTOUR MODELS [J].
KASS, M ;
WITKIN, A ;
TERZOPOULOS, D .
INTERNATIONAL JOURNAL OF COMPUTER VISION, 1987, 1 (04) :321-331
[9]
Mutual information for automated unwarping of rat brain autoradiographs [J].
Kim, B ;
Boes, JL ;
Frey, KA ;
Meyer, CR .
NEUROIMAGE, 1997, 5 (01) :31-40
[10]
Kossoff G, 1976, Ultrasound Med Biol, V2, P89, DOI 10.1016/0301-5629(76)90017-X