Transvenous removal of fibrin sheaths from tunneled hemodialysis catheters

被引:49
作者
Haskal, ZJ [1 ]
Leen, VH [1 ]
ThomasHawkins, C [1 ]
ShlanskyGoldberg, RD [1 ]
Baum, RA [1 ]
Soulen, MC [1 ]
机构
[1] HOSP UNIV PENN,MED CTR,DEPT MED,PHILADELPHIA,PA 19104
关键词
catheters and catheterization; central venous access; complications; dialysis; shunts;
D O I
10.1016/S1051-0443(96)70792-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: Long-term hemodialysis catheters are prone to significant dysfunction due to fibrin accumulation around their tips. The authors assessed the efficacy of transvenous snare removal of fibrin to prolong function of these catheters. PATIENTS AND METHODS: Twenty-four procedures were performed in 20 patients with tunneled hemodialysis central venous catheters. Technical success was gauged by venography and the ability to infuse and aspirate catheters. Durable efficacy was assessed by improvement in hemodialysis. RESULTS: Twenty-two of 24 procedures were performed successfully. In two cases residual material remained despite repeated stripping. Mean preprocedure hemodialysis blood-liters processed per hour increased from 15.1 to 19.1 L/h in the first dialysis session after stripping, and blood flow rates of 300 mL/min or greater were restored. By the fifth dialysis session after stripping, the blood-liters processed per hour dropped to 15.9 L/h as catheter flow rates returned to unacceptable levels. CONCLUSIONS: Percutaneous fibrin removal with a loop snare provides no durable benefit in improving function of failing hemodialysis catheters.
引用
收藏
页码:513 / 517
页数:5
相关论文
共 13 条
[1]   FIBRIN SHEATH ENTRAPMENT OF PERIPHERALLY INSERTED CENTRAL CATHETERS [J].
CARDELLA, JF ;
LUKENS, ML ;
FOX, PS .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1994, 5 (03) :439-442
[2]   NONINFECTIOUS COMPLICATIONS OF LONG-TERM CENTRAL VENOUS CATHETERS - RADIOLOGIC EVALUATION AND MANAGEMENT [J].
CASSIDY, FP ;
ZAJKO, AB ;
BRON, KM ;
REILLY, JJ ;
PEITZMAN, AB ;
STEED, DL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1987, 149 (04) :671-675
[3]   Fibrin sleeve stripping for salvage of failing hemodialysis catheters: Technique and initial results [J].
Crain, MR ;
Mewissen, MW ;
Ostrowski, GJ ;
PazFumagalli, R ;
Beres, RA ;
Wertz, RA .
RADIOLOGY, 1996, 198 (01) :41-44
[4]  
CRAIN MR, 1994, J AM SOC NEPHROL, V5, P412
[5]  
DAUGIRDAS JT, 1994, HDB DIALYSIS
[6]  
DUSZAK RL, 1995, RADIOLOGY, V197, P284
[7]   FIBRIN SHEET COVERING SUBCLAVIAN OR FEMORAL DIALYSIS CATHETERS [J].
HOMBROUCKX, R ;
DHALLUIN, F ;
BOGAERT, AM ;
LEROY, F ;
DEVOS, JY ;
LARNO, L .
ARTIFICIAL ORGANS, 1994, 18 (04) :322-324
[8]  
HOSHAL VL, 1971, ARCH SURG-CHICAGO, V102, P253
[9]   FUNCTIONAL RESTORATION OF OCCLUDED CENTRAL VENOUS CATHETERS - NEW INTERVENTIONAL TECHNIQUES [J].
KNELSON, MH ;
HUDSON, ER ;
SUHOCKI, PV ;
PAYNE, CS ;
SALLEE, DS ;
NEWMAN, GE .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1995, 6 (04) :623-627
[10]   DOES THE FIBRIN COAT AROUND A CENTRAL VENOUS CATHETER INFLUENCE CATHETER-RELATED SEPSIS [J].
LLOYD, DA ;
SHANBHOGUE, LKR ;
DOHERTY, PJ ;
SUNDERLAND, D ;
HART, CA ;
WILLIAMS, DF .
JOURNAL OF PEDIATRIC SURGERY, 1993, 28 (03) :345-349