Low-noise biomagnetic measurements with a multichannel dc-SQUID system at 77 K

被引:29
作者
Barthelmess, HJ
Halverscheid, M
Schiefenhövel, B
Heim, E
Schilling, M
Zimmermann, R
机构
[1] Univ Hamburg, Inst Angew Phys, D-20355 Hamburg, Germany
[2] Univ Hamburg, Zentrum Mikrostrukturforsch, D-20355 Hamburg, Germany
[3] Univ Klin Eppendorf, D-20246 Hamburg, Germany
关键词
dc-SQUIDs; YBa2Cu3O7; biomagnetism;
D O I
10.1109/77.919430
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Directly coupled de-SQUID magnetometers made from the high-temperature superconductor (HTS) YBa2Cu3O7 on symmetric 24 degrees and 30 degrees SrTiO3-bicrystals have been optimized for low-noise operation in static magnetic fields, These magnetometers are encapsulated, for reliable operation in liquid nitrogen. In our small glass cryostat, allowing even mobile applications, up to 24 magnetometers can be mounted in three levels for gradiometer arrangements. With first-order gradiometers an external noise suppression of more than a factor of 100 is achieved. In a magnetically shielded chamber we find a typical noise rootS(B)(77K, 10 Hz) of less than 50 fT/root Hz and low-noise magnetocardiograms are recorded. Due to the low noise of the de-SQUID sensors also electrically stimulated magnetoencephalography measurements were possible.
引用
收藏
页码:657 / 660
页数:4
相关论文
共 8 条
[1]  
BARTHELMESS HJ, 2000, IN PRESS P BIOMAG 20
[2]   High-T-c super conducting quantum interference devices with slots or holes: Low 1/f noise in ambient magnetic fields [J].
Dantsker, E ;
Tanaka, S ;
Clarke, J .
APPLIED PHYSICS LETTERS, 1997, 70 (15) :2037-2039
[3]   The development of a high-T-c magnetocardiography system for unshielded environment [J].
David, B ;
Dossel, O ;
Doormann, V ;
Eckart, R ;
Hoppe, W ;
Kruger, J ;
Laudan, H ;
Rabe, G .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1997, 7 (02) :3267-3270
[4]   Low-frequency noise and linearity of a YBa2Cu3O7 dc superconducting quantum interference device magnetometer in static magnetic fields [J].
Krey, S ;
Barthelmess, HJ ;
Schilling, M .
JOURNAL OF APPLIED PHYSICS, 1999, 86 (11) :6602-6604
[5]  
*NEUROSOFT INC, CURRYTM SOFTW DISPL
[6]   Biomagnetic measurements with an integrated YBa2Cu3O7 magnetometer in a hand-held cryostat [J].
Schilling, M ;
Krey, S ;
Scharnweber, R .
APPLIED PHYSICS LETTERS, 1996, 69 (18) :2749-2751
[7]  
SEIDEL P, 1997, IEEE T APPL SUPERCON, V7, P4077
[8]   Multichannel heart scanner based on high-Tc SQUIDs [J].
terBrake, HJM ;
Janssen, N ;
Flokstra, J ;
Veldhuis, D ;
Rogalla, H .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1997, 7 (02) :2545-2548