High accuracy measurements and Monte Carlo calculations of the relative efficiency curve of an HPGe detector from 433 to 2754 keV

被引:37
作者
Ludington, MA [1 ]
Helmer, RG
机构
[1] Albion Coll, Dept Phys, Albion, MI 49224 USA
[2] Idaho Natl Lab, Idaho Falls, ID 83415 USA
关键词
efficiency calibration; gamma rays; Monte Carlo; high-purity germanium detection system; Ge detectors;
D O I
10.1016/S0168-9002(99)01253-X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The relative full-energy peak efficiency curve of a 280-cm(3) coaxial HPGe detector was determined to an accuracy of about 0.1% over the energy range from 433 to 2754 keV from pairs of gamma-ray lines whose emission probabilities are very accurately known. The curve was extended from previous studies by the addition of lines from (108)mAg and Na-24. Ratios of full- energy peak efficiencies were also determined by Monte Carlo calculations using the CYLTRAN code. The measured ratios and the Monte Carlo values were in excellent agreement (0.03-0.19 %). The relative efficiency was defined by a pair of functions, one for 433-1332 and one for 614-2754 keV. Monte Carlo values were used to supplement the measured values, resulting in a more accurate fit. The stability of the relative efficiency curve for a 94-cm(3) HPGe detector was studied and it was found that there was no statistically significant change over a period of 3 years. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:506 / 521
页数:16
相关论文
共 12 条
[1]  
[Anonymous], 1986, NUCL SCI ENG, DOI DOI 10.13182/NSE86-A18182
[2]  
Hager R. S., 1968, Nuclear Data, Section A, V4, P1, DOI 10.1016/S0550-306X(68)80002-3
[3]   High-accuracy determination of the relative full energy peak efficiency curve of a coaxial HPGe detector in the high energy range 700-1300 keV [J].
Hawari, AI ;
Fleming, RF ;
Ludington, MA .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 398 (2-3) :276-286
[4]   HIGH-ACCURACY DETERMINATION OF THE SHAPE OF THE EFFICIENCY CURVE OF THE HPGE DETECTOR IN THE ENERGY-RANGE 900-KEV TO 1300-KEV [J].
HAWARI, AI ;
FLEMING, RF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 353 (1-3) :106-108
[5]  
HAWARI AI, 1995, THESIS U MICHIGAN AN
[6]   Highly accurate determination of γ-ray emission probabilities for 27Mg [J].
Hayashi, N ;
Miyahara, H ;
Hara, K ;
Mori, C ;
Fleming, RF ;
Dewaraja, YK ;
Ludington, MA .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 422 (1-3) :493-497
[7]   METHODOLOGY FOR THE MEASUREMENT OF BETA-DECAY INTENSITY DISTRIBUTIONS FROM THE ANALYSIS OF TOTAL ABSORPTION GAMMA-RAY SPECTRA [J].
HELMER, RG ;
PUTNAM, MH ;
GREENWOOD, RC ;
WILLMES, H .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 351 (2-3) :406-422
[8]  
LINDSTROM RM, 1993, P INT C NUCL AN METH
[9]  
MIYAHARA H, 1997, IN PRESS J INORG NUC, P93
[10]   RULES FOR SPIN AND PARITY ASSIGNMENTS BASED ON LOGFT VALUES [J].
RAMAN, S ;
GOVE, NB .
PHYSICAL REVIEW C, 1973, 7 (05) :1995-2009