Breath-by-breath measurement of carbon dioxide using a plastic film optical sensor

被引:74
作者
Mills, A
Lepre, A
Wild, L
机构
[1] Department of Chemistry, University of Wales Swansea, Swansea, SA28PP, Singleton Park
关键词
carbon dioxide; optical sensors; capnography;
D O I
10.1016/S0925-4005(96)02116-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An optical sensor has been developed for breath-by-breath gaseous CO2 analysis. The detector is based on a general formulation described in previous work where a phase-transfer agent, tetraoctyl ammonium hydroxide, is used to incorporate a hydrophilic pH-sensitive dye into a hydrophobic plastic film to create an effectively solid-state colorimetric sensor. In this work the formulation has been modified to yield a sensor which is capable of responding to clinically important levels of CO2 (0.1-5%) in less than 200 ms. This is comparable with the response of commercially available capnometers based on infrared sensing of CO2, which are currently widely used for clinical analysis. These films have good stability under the hot humid conditions involved in respiratory gas monitoring (37 degrees C gas temperature, 100% humidity) and, when stored in a freezer, show no deterioration in performance over an 80 day period.
引用
收藏
页码:419 / 425
页数:7
相关论文
共 21 条
[1]  
GAL TJ, 1990, CLIN MONITORING, P335
[2]  
GEDEON A, 1994, ANAESTHESIA, V49, P798
[3]  
HIGGINS DJ, 1993, ANESTH ANALG, V76, P683, DOI 10.1213/00000539-199303000-00075
[4]   SENSITIVITY OF A DISPOSABLE END-TIDAL CARBON-DIOXIDE DETECTOR [J].
JONES, BR ;
DORSEY, MJ .
JOURNAL OF CLINICAL MONITORING, 1991, 7 (03) :268-270
[5]  
LAUBER R, 1975, CAN J ANAESTHESIA, V42, P643
[6]  
LUBBERS DW, 1985, ACTA ANAESTH SCAND, V39, P37
[7]  
LUBBERS DW, 1983, ANAL CHEM S SER, V17, P609
[8]   EQUILIBRIUM STUDIES ON COLORIMETRIC PLASTIC FILM SENSORS FOR CARBON-DIOXIDE [J].
MILLS, A ;
CHANG, Q ;
MCMURRAY, N .
ANALYTICAL CHEMISTRY, 1992, 64 (13) :1383-1389
[9]   FLUORESCENCE PLASTIC THIN-FILM SENSOR FOR CARBON-DIOXIDE [J].
MILLS, A ;
CHANG, Q .
ANALYST, 1993, 118 (07) :839-843
[10]   MODELED DIFFUSION-CONTROLLED RESPONSE AND RECOVERY BEHAVIOR OF A NAKED OPTICAL FILM SENSOR WITH A HYPERBOLIC-TYPE RESPONSE TO ANALYTE CONCENTRATION [J].
MILLS, A ;
CHANG, Q .
ANALYST, 1992, 117 (09) :1461-1466