Alanine-polystyrene dosimeters prepared by injection moulding

被引:6
作者
Kojima, T
Kashiwazaki, S
Zhang, Y
机构
[1] HITACHI CABLE LTD, POWER SYST LAB, HITACHI, IBARAKI 31914, JAPAN
[2] NATL INST METROL, BEIJING 100013, PEOPLES R CHINA
关键词
D O I
10.1016/S0969-8043(97)00010-9
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The alanine-polystyrene dosimeters analysed by electron spin resonance (ESR) spectrometry were manufactured for large-scale use as a routine dosimeter in industrial radiation processing facilities by using an injection moulding method. Injection moulding enables us to achieve mass production of uniform dosimeters at a reduced cost. Their primary characteristics for dosimetry were studied in terms of their gamma-ray dose-response at radiation processing dose levels, precision, and effect of temperature during irradiation at an absorbed dose of 1 kGy. The precision of dose-response in the range 0.1-100 key is within +/- 2% at a 68% confidence level with and without correction for sample-to-sample mass scattering. The irradiation temperature coefficient for dosimeters irradiated to 1 kGy is estimated to be 0.25%/degrees C, by normalizing to those irradiated at 25 degrees C in the temperature range from 5 to 45 degrees C. The fading of the radiation-induced ESR signal at doses of 0.1-100 kGy at 25 degrees C and r.h. 40% is < 1% for at least 2 months after irradiation. (C) 1997 Elsevier Science Ltd.
引用
收藏
页码:965 / 968
页数:4
相关论文
共 15 条
[1]  
*ASTM, 1995, ANN BOOK ASTM STANDA, V1202, P855
[2]   USE OF ALANINE AS A SOLID DOSIMETER [J].
BRADSHAW, WW ;
SPETZLER, HA ;
CADENA, DG ;
CRAWFORD, GW .
RADIATION RESEARCH, 1962, 17 (01) :11-+
[3]   Orientation effects on ESR analysis of alanine-polymer dosimeters [J].
Kojima, T ;
Kashiwazaki, S ;
Tachibana, H ;
Tanaka, R ;
Desrosiers, MF ;
McLaughlin, WL .
APPLIED RADIATION AND ISOTOPES, 1995, 46 (12) :1407-1411
[4]   ALANINE ESR DOSIMETRY SYSTEM FOR ROUTINE USE IN RADIATION PROCESSING [J].
KOJIMA, T ;
HARUYAMA, Y ;
TACHIBANA, H ;
TANAKA, R ;
OKAMOTO, J ;
YAGI, K ;
TAMURA, N ;
HARA, H ;
KASHIWAZAKI, S .
RADIATION PHYSICS AND CHEMISTRY, 1993, 42 (4-6) :757-760
[5]   Irradiation and ESR analysis temperature dependence of the gamma-ray response of alanine-polystyrene dosimeters [J].
Kojima, T ;
Morishita, N ;
Itoh, H ;
Biramontri, S .
APPLIED RADIATION AND ISOTOPES, 1996, 47 (04) :457-459
[6]  
KOJIMA T, 1992, APPL RADIAT ISOTOPES, V43, P863, DOI 10.1016/0883-2889(92)90147-7
[7]  
KOJIMA T, 1989, APPL RADIAT ISOTOPES, V40, P851, DOI 10.1016/0883-2889(89)90006-3
[8]  
KOJIMA T, 1993, RADIOISOTOPES, V42, P607
[9]  
MAIER D, 1993, APPL RADIAT ISOTOPES, V44, P345
[10]  
McLaughlin W.L., 1989, DOSIMETRY RAD PROCES