Major stent deformation/pseudofracture of 7 Crown Endeavor/Micro Driver stent platform: incidence and causative factors

被引:33
作者
Pitney, Mark [1 ]
Pitney, Kate [2 ]
Jepson, Nigel [1 ]
Friedman, Daniel [1 ]
Nguyen-Dang, Tuan [1 ]
Matthews, Joseph [1 ]
Giles, Robert [1 ]
Taylor, David [3 ]
机构
[1] Prince Wales Hosp, Eastern Heart Clin, Sydney, NSW 2031, Australia
[2] Univ NSW, Sydney, NSW, Australia
[3] Bigge St Liverpool, Sydney SW Private, Sydney, NSW, Australia
关键词
stent designs; angioplasty; safety; cobalt-chromium stent; ELUTING CORONARY STENTS; TRIAL;
D O I
10.4244/EIJV7I2A41
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims: This is the initial report of stent deformation / pseudofracture of the 7 crown Endeavor / Micro Driver stent platform (2.25-2.75 mm), whereby the post-dilation balloon catches and causes major stent deformation angiographically appearing as a large stent fracture. We sought to determine frequency and cause. Methods and results: Analysis of 1,000 consecutive Endeavor/Micro Driver stents (7 crown) deployed at our institution. Bench testing was also performed by deploying 10 stents in 2.5 mm tubing and then attempting re-crossing with non-compliant balloons. Clinical results: There were 14 cases of major stent deformation/pseudofracture representing an incidence of 1.4% or 1.8% of 775 stents that were post-dilated. Of the 14 deformed stents, re-stenting was required in nine and MACE occurred in five (36%.). Benchtop results: Balloon "catch" was reproducible at an initial frequency of 9% and affected three of the 10 stents. With the use of provocative measures stent deformation occurred in 50% of test stents. Conclusions: Stent deformation/pseudofracture of the 7 crown Endeavor/Micro Driver platform occurred in 1.8% of cases that were post-dilated. Of these, 36% experience MACE. Deformation can be reproduced on the benchtop. Operators should be aware of the complication, and develop strategies to deal with it.
引用
收藏
页码:256 / 262
页数:7
相关论文
共 12 条
[1]  
Basalus MWZ, 2010, EUROINTERVENTION, V6, P141, DOI 10.4244/
[2]  
Dote K, 1991, J Cardiol, V21, P203
[3]  
Jin Xiaofeng, 2007, Int J Cardiol, V118, pe54, DOI 10.1016/j.ijcard.2006.12.064
[4]   Comparison of zotarolimus-eluting and sirolimus-eluting stents in patients with native coronary artery disease - A randomized controlled trial [J].
Kandzari, David E. ;
Leon, Martin B. ;
Popma, Jeffrey J. ;
Fitzgerald, Peter J. ;
O'Shaughnessy, Charles ;
Ball, Michael W. ;
Turco, Mark ;
Applegate, Robert J. ;
Gurbel, Paul A. ;
Midei, Mark G. ;
Badre, Sejal S. ;
Mauri, Laura ;
Thompson, Kweli P. ;
LeNarz, LeRoy A. ;
Kuntz, Richard E. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 48 (12) :2440-2447
[5]   A Randomized Comparison of the Endeavor Zotarolimus-Eluting Stent Versus the TAXUS Paclitaxel-Eluting Stent in De Novo Native Coronary Lesions 12-Month Outcomes From the ENDEAVOR IV Trial [J].
Leon, Martin B. ;
Mauri, Laura ;
Popma, Jeffrey J. ;
Cutlip, Donald E. ;
Nikolsky, Eugenia ;
O'Shaughnessy, Charles ;
Overlie, Paul A. ;
McLaurin, Brent T. ;
Solomon, Stuart L. ;
Douglas, John S., Jr. ;
Ball, Michael W. ;
Caputo, Ronald P. ;
Jain, Ash ;
Tolleson, Thaddeus R. ;
Reen, Bernard M., III ;
Kirtane, Ajay J. ;
Fitzgerald, Peter J. ;
Thompson, Kweli ;
Kandzari, David E. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (06) :543-554
[6]   Late thrombosis in drug-eluting coronary stents after discontinuation of antiplatelet therapy [J].
McFadden, EP ;
Stabile, E ;
Regar, E ;
Cheneau, E ;
Ong, ATL ;
Kinnaird, T ;
Suddath, WO ;
Weissman, NJ ;
Torguson, R ;
Kent, KM ;
Pichard, AD ;
Satler, LF ;
Waksman, R ;
Serruys, PW .
LANCET, 2004, 364 (9444) :1519-1521
[7]  
Meredith Ian T, 2005, EuroIntervention, V1, P157
[8]   Clinical and Angiographic Results With the Next-Generation Resolute Stent System A Prospective, Multicenter, First-in-Human Trial [J].
Meredith, Ian T. ;
Worthley, Stephen ;
Whitbourn, Robert ;
Walters, Darren L. ;
McClean, Dougal ;
Horrigan, Mark ;
Popma, Jeffrey J. ;
Cutlip, Donald E. ;
DePaoli, Ann ;
Negoita, Manuela ;
Fitzgerald, Peter J. .
JACC-CARDIOVASCULAR INTERVENTIONS, 2009, 2 (10) :977-985
[9]  
Park Seung-Jung, 2009, COMPARISON SIROLIMUS
[10]   A Qualitative and Quantitative Angiographic Analysis of Stent Fracture Late Following Sirolimus-Eluting Stent Implantation [J].
Popma, Jeffrey J. ;
Tiroch, Klaus ;
Almonacid, Alexandra ;
Cohen, Sidney ;
Kandzari, David E. ;
Leon, Martin B. .