Dynamic Nuclear Polarization Polarizer for Sterile Use Intent

被引:199
作者
Ardenkjaer-Larsen, Jan H. [1 ]
Leach, Andrew M. [2 ]
Clarke, Neil [3 ]
Urbahn, John [2 ]
Anderson, Denise [2 ]
Skloss, Timothy W. [4 ]
机构
[1] GE Healthcare, Brondby, Denmark
[2] Gen Elect, Global Res Ctr, Niskayuna, NY USA
[3] GE Healthcare, Abingdon, Oxon, England
[4] GE Healthcare, Waukesha, WI USA
关键词
hyperpolarization; dissolution-dynamic nuclear polarization; polarizer;
D O I
10.1002/nbm.1682
中图分类号
Q6 [生物物理学];
学科分类号
071011 [生物物理学];
摘要
A novel polarizer based on the dissolution-dynamic nuclear polarization (DNP) method has been designed, built and tested. The polarizer differs from those previously described by being designed with sterile use intent and being compatible with clinical use. The main features are: (1) an integral, disposable fluid path containing all pharmaceuticals constituting a sterile barrier, (2) a closed-cycle cryogenic system designed to eliminate consumption of liquid cryogens and (3) multi-sample polarization to increase throughput. The fluid path consists of a vial with the agent to be polarized, a pair of concentric inlet and outlet tubes connected to a syringe with dissolution medium and a receiver, respectively. The fluid path can operate at up to 400 K and 2.0 MPa and generates volumes as high as 100 mL. An inline filter removes the amount of electron paramagnetic agent in the final product by more than 100-fold in the case of [1-(13)C]pyruvate. The system uses a sorption pump in conjunction with a conventional cryocooler. The system operates through cycles of pumping to low temperature and regeneration of the sorption pump. The magnet accommodates four samples at the same time. A temperature of less than 1 K was achieved for 68 h (no sample heat loads) with a liquid helium volume of 2.4 L. The regeneration of the liquid helium could be achieved in less than 10 h, and the transition to cold (< 1.2 K) was achieved in less than 90 min. A solid state polarization of 36 +/- 4% for [1-(13)C]pyruvic acid was obtained with only 10 mW of microwave power. The loading of a sample adds less than 50 J of heat to the helium bath by introducing the sample over 15 min. The heat load imposed on the helium bath during dissolution was less than 70 J. The measured liquid state polarization was 18 +/- 2%. Copyright (C) 2011 John Wiley & Sons, Ltd.
引用
收藏
页码:927 / 932
页数:6
相关论文
共 8 条
[1]
Increase in signal-to-noise ratio of &gt;10,000 times in liquid-state NMR [J].
Ardenkjaer-Larsen, JH ;
Fridlund, B ;
Gram, A ;
Hansson, G ;
Hansson, L ;
Lerche, MH ;
Servin, R ;
Thaning, M ;
Golman, K .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2003, 100 (18) :10158-10163
[2]
ARDENKJAERLARSE.JH, 2010, MOL IMAGING PRINCIPL
[3]
Imaging and 'omic' methods for the molecular diagnosis of cancer [J].
Bohndiek, Sarah E. ;
Brindle, Kevin M. .
EXPERT REVIEW OF MOLECULAR DIAGNOSTICS, 2010, 10 (04) :417-434
[4]
Design and performance of a DNP prepolarizer coupled to a rodent MRI scanner [J].
Comment, A. ;
van den Brandt, B. ;
Uffmann, K. ;
Kurdzesau, F. ;
Jannin, S. ;
Konter, J. A. ;
Hautle, P. ;
Wenckebach, W. TH. ;
Gruetter, R. ;
van der Klink, J. J. .
CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2007, 31B (04) :255-269
[5]
Biomedical applications of hyperpolarized 13C magnetic resonance imaging [J].
Gallagher, F. A. ;
Kettunen, M. I. ;
Brindle, K. M. .
PROGRESS IN NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY, 2009, 55 (04) :285-295
[6]
Jain J., 2008, P ASME 2008 HEAT TRA, P559, DOI [10.1115/HT2008-56148, DOI 10.1115/HT2008-56148]
[7]
*OXF INSTR, HYP PROD
[8]
URBAHN JA, INT CRYOG ENG C 23 I