Noninvasive cardiac output by partial CO2 rebreathing after severe chest trauma

被引:25
作者
Maxwell, RA
Gibson, JB
Slade, JB
Fabian, TC
Proctor, KG
机构
[1] Univ Tennessee, Ctr Hlth Sci, Dept Surg, Memphis, TN 38163 USA
[2] Univ Tennessee, Ctr Hlth Sci, Dept Anesthesiol, Memphis, TN 38163 USA
[3] Univ Tennessee, Ctr Hlth Sci, Dept Physiol, Memphis, TN 38163 USA
[4] Univ Tennessee, Erlanger Med Ctr, Dept Surg, Chattanooga, TN USA
来源
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE | 2001年 / 51卷 / 05期
关键词
noninvasive cardiac output monitoring; Fick principle; pulmonary contusion; swine;
D O I
10.1097/00005373-200111000-00005
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: In multiple trauma patients, early continuous cardiac output (CCO) monitoring is frequently desired but is difficult to routinely employ in most emergency departments because it requires invasive procedures. Recently, a noninvasive cardiac output (NICO) technique based on the Fick principle and partial CO2 rebreathing has shown promise under a variety of conditions. Since this method has not been tested after lung damage, we evaluated its utility in a clinically relevant model. Methods: Anesthetized, ventilated swine (n = 11, 35-45 kg) received a unilateral blunt trauma via a captive bolt gun followed by a 25% hemorrhage. After 60 min of shock, crystalloid resuscitation was given as needed to maintain heart rate < 100 beats/min and mean arterial pressure > 70 min Hg. Standard CCO by thermodilution (Baxter Vigilance, Irvine, CA) was compared with NICO (Novametrix Medical Systems Inc., Wallingford, CT) for 8 hr. Results: The severity of the injury is reflected by seven deaths (average survival time = 4.25 hr). Trauma increased dead space ventilation (19%), airway resistance (30%), and lactate (3.2 mmol/L), and decreased dynamic compliance (48%) and Pao(2)/FIO2 (54%). In these extreme conditions, the time course and magnitude of change of CCO and NICO were superimposed. Bland-Altman analysis reveal a bias and precision of 0.01 +/- 0.69 liters/min. The linear relationship between individual CCO and NICO values was significant (p < 0.0001) and was described by the equation NICO = (0.74 +/- 0.1)CCO + (0.65 +/- 0.16 liters/min) but the correlation coefficient (r(2) = 0.541) was relatively low. The cause for the low correlation could not be attributed to increased pulmonary shunt, venous desaturation, anemia, hypercapnia, increased dead space ventilation, or hyperlactacidemia. Conclusion: NICO correlated with thermodilution CCO, but underestimated this standard by 26% in extreme laboratory conditions of trauma-induced cardiopulmonary dysfunction; 95% of the NICO values fall within 1.38 liters/min of CCO; and with further improvements, NICO may be useful in multiple trauma patients requiring emergency intubation during initial assessment and workup.
引用
收藏
页码:849 / 853
页数:5
相关论文
共 13 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   NONINVASIVE MEASUREMENT OF CARDIAC-OUTPUT USING PARTIAL CO2 REBREATHING [J].
CAPEK, JM ;
ROY, RJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (09) :653-661
[3]   Partial carbon dioxide rebreathing: A reliable technique for noninvasive measurement of nonshunted pulmonary capillary blood flow [J].
deAbreu, MG ;
Quintel, M ;
Ragaller, M ;
Albrecht, DM .
CRITICAL CARE MEDICINE, 1997, 25 (04) :675-683
[4]  
Epstein RH, 1999, ANESTHESIOLOGY, V91, pU271
[5]   A NEW METHOD FOR NON-INVASIVE BEDSIDE DETERMINATION OF PULMONARY BLOOD-FLOW [J].
GEDEON, A ;
FORSLUND, L ;
HEDENSTIERNA, G ;
ROMANO, E .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1980, 18 (04) :411-418
[6]   Clinical evaluation of a new noninvasive method of cardiac output measurement - Preliminary results in CABG patients [J].
Guzzi, L ;
Jaffe, MB ;
Orr, JA .
ANESTHESIOLOGY, 1998, 89 (3A) :U409-U409
[7]  
HARYADI DG, 1998, ANESTHESIOLOGY, V89, pA534
[8]   End-tidal carbon dioxide as a noninvasive indicator of cardiac index during circulatory shock [J].
Jin, XH ;
Weil, MH ;
Tang, WC ;
Povoas, H ;
Pernat, A ;
Xie, JL ;
Bisera, J .
CRITICAL CARE MEDICINE, 2000, 28 (07) :2415-2419
[9]  
JOHNSON KB, 1998, ANESTHESIOLOGY, V89, pA535
[10]  
Loeb Robert G., 1999, Anesthesiology (Hagerstown), V91, pA474