Instrumentation developments for X-ray linear and circular dichroism at the ESRF beamline ID12A

被引:72
作者
Goulon, J
Rogalev, A
Gauthier, C
Goulon-Ginet, C
Paste, S
Signorato, R
Neumann, C
Varga, L
Malgrange, C
机构
[1] European Synchrotron Radiat Facil, F-38043 Grenoble, France
[2] Universite Joseph Fourier, Fac Pharm, F-38041 Grenoble, France
[3] Univ Paris 06, Lab Mineral Cristallog, F-75252 Paris, France
[4] Univ Paris 07, F-75252 Paris 05, France
关键词
X-ray magnetic circular dichroism (X-MCD); X-ray magnetic linear dichroism (X-MLD); X-ray natural circular dichroism (X-NCD); X-ray polarimetry;
D O I
10.1107/S090904959701409X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper reports on the performance of the instrumentation developed for the ESRF beamline ID12A, which is dedicated to spectroscopic applications requiring full control of the polarization at energies greater than or equal to 2.0 keV. Emphasis is placed on the characterization of various optical components of the beamline and on problems associated with either the control or the conversion of the polarization state. A few examples have been selected to illustrate what sort of new spectroscopic information has already been obtained at the beamline. These include the comparison of X-ray magnetic dichroism spectra recorded with linear or circular polarization and the very first detection of X-ray natural circular dichroism in single crystals known to exhibit a very large non-linear susceptibility at optical wavelengths.
引用
收藏
页码:232 / 238
页数:7
相关论文
共 24 条