Fatigue behaviors of HP-Mg, Mg-Ca and Mg-Zn-Ca biodegradable metals in air and simulated body fluid

被引:102
作者
Bian, Dong [1 ]
Zhou, Weirui [2 ]
Liu, Yang [1 ]
Li, Nan [1 ]
Zheng, Yufeng [1 ,2 ]
Sun, Zhili [3 ]
机构
[1] Peking Univ, Dept Mat Sci & Engn, Coll Engn, Beijing 100871, Peoples R China
[2] Peking Univ, Ctr Biomed Mat & Tissue Engn, Acad Adv Interdisciplinary Studies, Beijing 100871, Peoples R China
[3] Peking Univ, Coll Engn, State Key Lab Turbulence & Complex Syst, Beijing 100871, Peoples R China
基金
中国国家自然科学基金;
关键词
Biomedical magnesium alloy; Fatigue; Corrosion fatigue; Simulated body fluid; MAGNESIUM ALLOYS; CORROSION; PERFORMANCE; IMPLANTS; FRACTURE; FAILURE; SAFETY; AZ61; MULTICENTER; FIXATION;
D O I
10.1016/j.actbio.2016.05.031
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
The dynamic loading in human body, along with the corrosive body fluid, presents a great challenge for the practical use of biodegradable magnesium implants. In this study, a high purity magnesium (99.99 wt.%) and two typical promising biodegradable magnesium alloys (binary Mg-1 Ca and ternary Mg-2Zn-0.2Ca) were chosen as the experimental materials. Their dynamic mechanical performances were comparatively evaluated by carrying out fatigue tests in air and in simulated body fluid (SBF). The fatigue strengths of HP-Mg, Mg-1Ca and Mg-2Zn-0.2Ca were all around 90 MPa in air, however, they decreased to 52 MPa, 70 MPa and 68 MPa in SBF at 4 x 10(6) cycles, respectively. The fatigue cracks initiated from the microstructural defects when tested in air, but nucleated from surface corrosion pits when tested in SBF. Cyclic loading significantly increased the corrosion rates of all the experimental materials compared to that in static SBF. Moreover, based on our findings, the fatigue failure processes and interactions between material, corrosion and cyclic loading were systematically discussed.
引用
收藏
页码:351 / 360
页数:10
相关论文
共 59 条
[1]
Neuropathology of aluminum toxicity in rats (glutamate and GABA impairment) [J].
Abd El-Rahman, SS .
PHARMACOLOGICAL RESEARCH, 2003, 47 (03) :189-194
[2]
In-body corrosion fatigue failure of a stainless steel orthopaedic implant with a rare collection of different damage mechanisms [J].
Amel-Farzad, H. ;
Peivandi, M. T. ;
Yusof-Sani, S. M. R. .
ENGINEERING FAILURE ANALYSIS, 2007, 14 (07) :1205-1217
[3]
[Anonymous], ACTA METALL SIN
[4]
[Anonymous], 2004, ASTM E8 04
[5]
[Anonymous], 2004, G3172 ASTM
[6]
[Anonymous], ASTME46696
[7]
[Anonymous], RUSS METALL
[8]
Corrosion mechanism applicable to biodegradable magnesium implants [J].
Atrens, Andrej ;
Liu, Ming ;
Abidin, Nor Ishida Zainal .
MATERIALS SCIENCE AND ENGINEERING B-ADVANCED FUNCTIONAL SOLID-STATE MATERIALS, 2011, 176 (20) :1609-1636
[9]
Failure analysis of a commercially pure titanium plate for osteosynthesis [J].
Azevedo, CRF .
ENGINEERING FAILURE ANALYSIS, 2003, 10 (02) :153-164
[10]
Corrosion fatigue behavior of extruded magnesium alloy AZ61 under three different corrosive environments [J].
Bhuiyan, Md. Shahnewaz ;
Mutoh, Yoshiharu ;
Murai, Tsutomu ;
Iwakami, Shinpei .
INTERNATIONAL JOURNAL OF FATIGUE, 2008, 30 (10-11) :1756-1765