Transcatheter closure of persistent arterial ducts with different types of coils

被引:11
作者
Cheung, YF [1 ]
Leung, MP [1 ]
Chau, RT [1 ]
机构
[1] Univ Hong Kong, Grantham Hosp, Dept Pediat, Div Pediat Cardiol, Aberdeen, Hong Kong, Peoples R China
关键词
D O I
10.1067/mhj.2001.111263
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Different types of coils have been designed for transcatheter closure of persistent arterial duels. We ccm pared the efficacy and safety of three types of coils: Gianturco coils (Cook), Cook detachable coils (Cook), and Duct Occlud devices (pfm). Methods Sixty-three patients underwent coil occlusion of arterial duels between April 1995 and July 2000. The mean age and weight were 4.8 +/- 3.4 years and 16.5 +/- 7.6 kg, respectively. The results and complications of ductal occlusion among the three types of coils were compared. Kaplan-Meier analysis was used to assess reduction in the prevalence of residual shunt with time, and multiple regression analysis was performed to identify predictors of complete occlusion. Results Coil occlusion of persistent arterial ducts that measured 2.2 +/- 0.8 mm was feasible in 90% (57/63) of patients. Gianturco coils were used in 29, Duct Occlud devices in 16, and Cook detachable coils in 12 patients. The prevalence of residual shunt at 24 hours, 6 months, 12 months, and 24 months was 42%, 20%, 18%, and 14%, respectively. The reduction in prevalence of residual shunt with time tended to be greater when Gianturco coils were used (P = .067). Logistic regression identified the use of Gianturco coils to be a significant predictor of complete ductal occlusion on Follow-up (P = .04). Pull-through of coils occurred in 4.8% (3/63) and coil embolization in 6.3% (4/63). There was no association between the type of coil and the risk of embolization (P = 1.00). Conclusions Transcatheter occlusion of small persistent arterial ducts with coils is safe and effective. There is no advantage of detachable coils (Cook detachable coils and Duct Occlud devices) over nondetachable Gianturco coils in reducing the risk of embolization. Our Findings are in favor of the inexpensive, but more effective, Gianturco coils for occluding smell arterial ducts of 3 mm or less.
引用
收藏
页码:87 / 91
页数:5
相关论文
共 19 条
[1]   PERCUTANEOUS CLOSURE OF THE SMALL (LESS-THAN-2.5 MM) PATENT DUCTUS-ARTERIOSUS USING COIL EMBOLIZATION [J].
CAMBIER, PA ;
KIRBY, WC ;
WORTHAM, DC ;
MOORE, JW .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (08) :815-816
[2]   Pulmonary artery size and flow disturbances after patent ductus arteriosus coil occlusion [J].
Carey, LM ;
Vermilion, RP ;
Shim, D ;
Lloyd, TT ;
Beekman, RH ;
Ludomirsky, A .
AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (11) :1307-&
[3]   Early implantation of multiple spring coils for severe haemolysis after incomplete transcatheter occlusion of persistent arterial duct [J].
Cheung, YF ;
Leung, MP ;
Chau, KT .
HEART, 1997, 77 (05) :477-478
[4]  
GRAY DT, 1993, NEW ENGL J MED, V329, P1570
[5]   RESULTS OF ANTEROGRADE TRANSCATHETER CLOSURE OF PATENT DUCTUS-ARTERIOSUS USING SINGLE OR MULTIPLE GIANTURCO COILS [J].
HIJAZI, ZM ;
GEGGEL, RL .
AMERICAN JOURNAL OF CARDIOLOGY, 1994, 74 (09) :925-929
[6]  
HIJAZI ZM, 1996, HEART, V76, P535
[7]  
LE TP, 1993, CARDIOL YOUNG, V3, P1
[8]   TRANSCATHETER OCCLUSION OF PATENT DUCTUS-ARTERIOSUS WITH GIANTURCO COILS [J].
LLOYD, TR ;
FEDDERLY, R ;
MENDELSOHN, AM ;
SANDHU, SK ;
BEEKMAN, RH .
CIRCULATION, 1993, 88 (04) :1412-1420
[9]   Catheter closure of moderate- to large-sized patent ductus arteriosus using the new Amplatzer duct occluder: Immediate and short-term results [J].
Masura, J ;
Walsh, KP ;
Thanopoulous, B ;
Chan, C ;
Bass, J ;
Goussous, Y ;
Gavora, P ;
Hijazi, ZM .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (04) :878-882
[10]   PERCUTANEOUS CLOSURE OF THE SMALL PATENT DUCTUS-ARTERIOSUS USING OCCLUDING SPRING COILS [J].
MOORE, JW ;
GEORGE, L ;
KIRKPATRICK, SE ;
MATHEWSON, JW ;
SPICER, RL ;
UZARK, K ;
ROTHMAN, A ;
CAMBIER, PA ;
SLACK, MC ;
KIRBY, WC .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (03) :759-765